Fluke 1552A-12 User Manual - Page 14
A/1552A Ex, application. See
View all Fluke 1552A-12 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 14 highlights
1551A/1552A Ex Users Manual The minimum immersion depth for the Probe is 3.5 inches in a stirred fluid heat source. It is recommended that you make sure of the minimum immersion depth for the specific application. See Figure 2. • The maximum immersion depth for the 1551A Ex is 2 inches less than the total length. Make sure the last 2 inches of Probe, nearest to the internal electronics of the device, are not put into the heat source. • The maximum immersion depth for the 1552A Ex is 3 inches less than the total length. Make sure the last 3 inches of the Probe, nearest to the internal electronics of the device, are not put into the heat source. A 1552A Ex: Maximum immersion: 3 inches less than total probe length. B 1551A Ex: Maximum immersion: 2 inches less than total probe length. Figure 2. Immersion Depth 6 gjo002.eps