Fluke 192C FE 192,196,199 C Users Manual - Page 70
Triggering on Narrow Glitches, look at all the stored glitches.
View all Fluke 192C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 70 highlights
Fluke 192B - 196B/C - 199B/C Users Manual To set the pulse width to 5 ms, do the following: 7 Enable the arrow keys to adjust the pulse width. 8 Select 5 ms. All narrow positive pulses shorter than 5 ms are now displayed on the screen. (See Figure 33.) Tip The test tool stores all triggered screens in the replay memory. For example, if you setup your triggering for glitches, you can capture 100 glitches with time stamps. Use the REPLAY key to look at all the stored glitches. Figure 33. Triggering on Narrow Glitches 62
Fluke 192B - 196B/C - 199B/C
Users Manual
62
To set the pulse width to 5 ms, do the following:
7
Enable the arrow keys to adjust
the pulse width.
8
Select 5 ms.
All narrow positive pulses shorter than 5 ms are now
displayed on the screen. (See Figure 33.)
Tip
The test tool stores all triggered screens in the
replay memory. For example, if you setup your
triggering for glitches, you can capture 100
glitches with time stamps. Use the
REPLAY
key to
look at all the stored glitches.
Figure 33. Triggering on Narrow Glitches