Fluke 355 Fluke 353 & 355 Users Manual - Page 24
Voltage Measurement 355 only 10 Hz to 100 Hz, Ohms Measurement 355 only, Range, Resolution, Accuracy
View all Fluke 355 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 24 highlights
353/355 Users Manual Voltage Measurement (355 only) 10 Hz to 100 Hz 600 and 1000 V ranges have 10 % over range to 660 and 1100 V respectively. Range 4 V Resolution 1 mV Accuracy 1 % rdg + 10 digits Trigger Level for Hz Filter OFF 0.050 V 40 V 10 mV 1 % rdg + 5 digits 0.25 V 400 V 100 mV 1 % rdg + 5 digits 6 V 600 V AC RMS 1 V 1 % rdg + 5 digits 6 V 1000 V DC 1 V 1 % rdg + 5 digits N/A Voltage Measurement (355 only) 100.1 Hz to 1 kHz 600 and 1000 V ranges have 10 % over range to 660 and 1100 V Range 4 V Resolution 1 mV Accuracy 3 % rdg + 10 digits 40 V 10 mV 3 % rdg + 5 digits 400 V 100 mV 3 % rdg + 5 digits 600 V AC RMS 1 V 3 % rdg + 5 digits Ohms Measurement (355 only) Range 400 Ω 4 k Ω 40 k Ω 400 k Ω Resolution 0.1 Ω 1 Ω 10 Ω 100 Ω Accuracy 1.5 % + 5 digits 1.5 % + 5 digits 1.5 % + 5 digits 1.5 % + 5 digits Trigger Level for Hz Filter ON 0.050 V 0.25 V 6 V 6 V N/A 16