Warning
View all Invacare RPL450-1 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 16 highlights
5 ASSEMBLY Assembling the Mast to the Base 1. If locking-type casters are on the patient lift, lock them. 2. Remove the shoulder bolt, nut and washer, that secures the mast in the U-shape cut-out of the base. 3. Position the mast in an upright position and place the mast into the U-shaped cut-out of the base. 4. Insert shoulder bolt with washers through the base and mast. 5. Secure with nut. ƽ WARNING The mast may be removed from the base for storage or transporting. Each time the mast is removed and returned to the base, the mast MUST be properly secured to the base assembly. Step here to LOCK. Step here to UNLOCK. Locking Lever Mast Base Washers Shoulder Bolt Nut Mast U-shaped Cutout Shoulder Bolt Washers Nut Manual/Electric Portable Patient Lift 16 Part No. 1078987