Invacare RPL600-1 Owners Manual - Page 48

Maintaining the Base Adjustment, Replacing Casters/Forks, Replacing Rear Casters

Page 48 highlights

10 MAINTENANCE 10.7 Maintaining the Base Adjustment The base adjustment should not require any attention other than: 1. Check the squareness of the legs when in the closed position. 2. Place a square on the inside of the legs and base to determine the 90° alignment. 3. Adjust the linkage rods until 90° alignment is achieved. Square Linkage Rods 90° Linkage Rod 10.8 Replacing Casters/Forks Replacing Rear Casters 1. Place the lift on its side. 2. Remove the bolt and locknut that secure the existing rear caster to the fork.  The bushing will be loose and may fall out of the caster. Existing bushing will be reused. Examine and replace if worn. 3. Install the new/existing bushing into the new rear caster. 4. Line up the mounting holes in the new rear caster and the fork. 5. Install the bolt through the fork and new rear caster and tighten securely with the locknut. Manual/Electric Portable Patient Lift 48 Locknut Rear Caster Bushing Base Fork Bolt Part No. 1078987

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60

10 MAINTENANCE
Manual/Electric Portable Patient Lift
48
Part No. 1078987
10.7
Maintaining the Base Adjustment
The base adjustment should not require any attention other than:
1.
Check the squareness of the legs when in the closed position.
2.
Place a square on the inside of the legs and base to determine the
90° alignment.
3.
Adjust the linkage rods until 90° alignment is achieved.
10.8
Replacing Casters/Forks
Replacing Rear Casters
1.
Place the lift on its side.
2.
Remove the bolt and locknut that secure the existing rear caster to
the fork.
3.
Install the new/existing bushing into the new rear caster.
4.
Line up the mounting holes in the new rear caster and the fork.
5.
Install the bolt through the fork and new rear caster and tighten
securely with the locknut.
90°
Square
Linkage Rods
Linkage Rod
The bushing will be loose and may fall out of the caster.
Existing bushing will be reused. Examine and replace if
worn.
Base
Fork
Bolt
Bushing
Rear Caster
Locknut