LiftMaster GH GH LOGIC VERSION 1 Manual - Page 14

OPTIONAL CONTROL SETTINGS, Set Timer to Close NOTE: Requires P/N 1A4811 CPSII

Page 14 highlights

OPTIONAL CONTROL SETTINGS Set / Reset Maximum Run Timer Begin with door in closed position. Set dip switch to max. run timer mode. Press control station open button to operate door from closed to full open position without stopping. Set dip switch to desired operating mode (B2, C2, D1, E2, T, TS). Set max. run timer 1234 OFF Adjustable Mid Stop Set: Begin with door in closed position. Set dip switch to adj. mid stop mode. Press control station open button to operate door from closed to mid stop position and stop with control station stop button. Set dip switch to desired operating mode (B2, C2, D1, E2, T, TS). Clear: Begin with door in closed positon. Set operator in set mid stop mode. Press control station open button. Allow the door to run to the open limit. Set the dip switch to desired operating mode (B2, C2, D1, E2, T, TS). Set adj. mid stop 1234 OFF Set Timer to Close (NOTE: Requires P/N 1A4811 CPSII Option Board with Timer to Close Function.) Set dip switch to timer to close mode. Momentarily press control station open button to set timer duration in 5 second increments. (Red diagnostic L.E.D. will flash to indicate the entry of each 5 second increment into memory). To re-set timer memory to zero, press control station close button. Set dip switch to (T or TS) operating mode after timer is programmed. Set timer to close 1234 OFF Diagnostic Mode Set dip switch to diagnostic mode. Flashing red diagnostic L.E.D. indicates proper microprocessor function. If the diagnostic L.E.D. does not light, the control logic board requires replacement. 1234 Diagnostic mode OFF 14

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20

Set / Reset Maximum Run Timer
Begin with door in closed position. Set dip switch to max. run timer
mode. Press control station open button to operate door from closed to
full open position without stopping. Set dip switch to desired operating
mode (B2, C2, D1, E2, T, TS).
Adjustable Mid Stop
Set:
Begin with door in closed position. Set dip switch to adj. mid stop
mode. Press control station open button to operate door from closed to
mid stop position and stop with control station stop button. Set dip switch
to desired operating mode (B2, C2, D1, E2, T, TS).
Clear:
Begin with door in closed positon.
Set operator in set mid stop
mode.
Press control station open button.
Allow the door to run to the
open limit.
Set the dip switch to desired operating mode (B2, C2, D1,
E2, T, TS).
Set Timer to Close (NOTE: Requires P/N 1A4811 CPSII
Option Board with Timer to Close Function.)
Set dip switch to timer to close mode. Momentarily press control station
open button to set timer duration in 5 second increments. (Red
diagnostic L.E.D. will flash to indicate the entry of each 5 second
increment into memory). To re-set timer memory to zero, press control
station close button. Set dip switch to (T or TS) operating mode after
timer is programmed.
Diagnostic Mode
Set dip switch to diagnostic mode. Flashing red diagnostic L.E.D.
indicates proper microprocessor function. If the diagnostic L.E.D. does
not light, the control logic board requires replacement.
Set
max.
run
timer
1
2
3
OFF
4
OPTIONAL CONTROL SETTINGS
14
Set
adj.
mid
stop
1
2
3
OFF
4
Set
timer
to close
1
2
3
OFF
4
Diagnostic
mode
1
2
3
OFF
4