Sanyo MPR-1013R Instruction Manual - Page 13

Installation

Page 13 highlights

INSTALLATION Installation 1. Remove the packaging materials and tapes Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the outside panels are dirty, clean them with a neutral detergent and wipe it off with a soft cloth washed in clean water. 2. Adjust the leveling foot Extend the leveling feet by rotating them counterclockwise until they make contact with the floor. See Fig. 1. Ensure the unit is level. Leveling foot Fig. 1 3. Fix the unit Two fixtures are attached to the rear of the frame. Fix the frame to the wall by attaching a rope or chain between the wall and the fixtures. 1) Reattach the fixtures as shown in Fig. 2. 2) If holes can be drilled in the wall, drill a 10.5mm diameter hole and fix the unit by using the special bolt-nut supplied. This bolt-nut can only be used on a concrete wall. See Fig. 3. 4. Ground (earth) Bolt (provided) Fig. 2 45 mm 10.5 mm Fig. 3 WARNING Use a power supply outlet with ground (earth) to prevent electric shock. If the power supply outlet is not grounded, it will be necessary to install a ground by qualified engineers. Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such grounding may cause electric shock in the case of an incomplete circuit. 12

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28

12
INSTALLATION
Installation
1. Remove the packaging materials and tapes
Remove all transportation packaging materials and
tapes.
Open the doors and ventilate the unit.
If the
outside panels are dirty, clean them with a neutral
detergent and wipe it off with a soft cloth washed in
clean water.
2. Adjust the leveling foot
Extend the leveling feet by rotating them
counterclockwise until they make contact with the floor.
See Fig. 1.
Ensure the unit is level.
3. Fix the unit
Two fixtures are attached to the rear of the frame.
Fix
the frame to the wall by attaching a rope or chain
between the wall and the fixtures.
1) Reattach the fixtures as shown in Fig. 2.
2) If holes can be drilled in the wall, drill a 10.5mm
diameter hole and fix the unit by using the special
bolt-nut supplied.
This bolt-nut can only be used on a
concrete wall.
See Fig. 3.
4. Ground (earth)
WARNING
Use a power supply outlet with ground (earth)
to prevent electric shock. If the power supply outlet is
not grounded, it will be necessary to install a ground by qualified engineers.
Never ground the unit through a gas pipe, water main, telephone line or lightning rod.
Such
grounding may cause electric shock in the case of an incomplete circuit.
Fig. 1
Fig. 2
Fig. 3
Bolt (provided)
45 mm
10.5 mm
Leveling foot