Sanyo mpr-214f Instruction Manual - Page 28
points to avoid the crush of the tube when bending the capillary tube.
View all Sanyo mpr-214f manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 28 highlights
INSTALLATION OF TEMPERATURE RECORDER 4. Remove the caps (both outside and inside) covering the access port for freezer and take out the thermal insulation in the port. Pass the recorder sensor to the chamber through the port. Then replace the thermal insulation and caps (both outside and inside). (Fig.13) Note: Make a cut on the caps for access port to pass the capillary tube as shown in the figure below. Fig. 13 5. Fix the recorder sensor to the attachment hole on the left side of the freezer by using the middle size clips enclosed with the recorder fitting (MPR-S30: optional component). (Fig. 14) 6. Arrange the capillary tube so that it cannot contact the shelf. Care should be taken for the following points to avoid the crush of the tube when bending the capillary tube. ɾDo not bend the tube of 30 mm from the edge of the sensor. ɾDo not bend and extend the tube repeatedly at the same point. ɾThe bending should be R10ʙR15. Fig. 14 27