Campbell Scientific HMP60 HMP60 Temperature and Relative Humidity Probe - Page 15

Campbell Scientific HMP60 Manual

Page 15 highlights

HMP60 Temperature and Relative Humidity Probe Probe Body Diameter: Filter: Filter Diameter: Housing Body Material: Filter Cap Material: Classification: Power Consumption: Supply Voltage: Settling Time after power is switched on: Output Signal Range: 1.2 cm (0.47 in) 0.2 μm Teflon membrane 1.2 cm (0.47 in) AISI 316 stainless steel Chrome-coated ABS plastic IP65 1 mA typical; 5 mA maximum 5 to 28 Vdc 1 second 0 to 1 Vdc 6.1 Temperature Sensor Sensor: Temperature Measurement Range: Temperature Accuracy: 1000 Ω PRT, DIN 43760B -40° to +60°C ±0.6°C (-40° to +60°C) 6.2 Relative Humidity Sensor Sensor: Relative Humidity Measurement Range: Accuracy at 0° to +40°C: INTERCAP® 0 to 100% non-condensing ±3% RH (0 to 90% Relative Humidity) ±5% RH (90 to 100% Relative Humidity) ±5% RH (0 to 90% Relative Humidity) ±7% RH (90 to 100% Relative Humidity) Accuracy at -40° to 0°C and +40° to +60°C: CAUTION The black outer jacket of the cable is Santoprene® rubber. This compound was chosen for its resistance to temperature extremes, moisture, and UV degradation. However, this jacket will support combustion in air. It is rated as slow burning when tested according to U.L. 94 H.B. and will pass FMVSS302. Local fire codes may preclude its use inside buildings. 9

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26

HMP60 Temperature and Relative Humidity Probe
Probe Body Diameter:
1.2 cm (0.47 in)
Filter:
0.2
μ
m Teflon membrane
Filter Diameter:
1.2 cm (0.47 in)
Housing
Body Material:
AISI 316 stainless steel
Filter Cap Material:
Chrome-coated ABS plastic
Classification:
IP65
Power Consumption:
1 mA typical; 5 mA maximum
Supply Voltage:
5 to 28 Vdc
Settling Time after
power is switched on:
1 second
Output Signal Range:
0 to 1 Vdc
6.1
Temperature Sensor
Sensor:
1000
Ω
PRT, DIN 43760B
Temperature
Measurement Range:
-40
°
to +60
°
C
Temperature Accuracy:
±0.6°C (-40° to +60°C)
6.2
Relative Humidity Sensor
Sensor:
INTERCAP
®
Relative Humidity
Measurement Range:
0 to 100% non-condensing
Accuracy at 0
°
to +40°C:
±
3% RH (0 to 90% Relative Humidity)
±
5% RH (90 to 100% Relative Humidity)
Accuracy at -40° to 0°C
and +40° to +60°C:
±
5% RH (0 to 90% Relative Humidity)
±
7% RH (90 to 100% Relative Humidity)
The black outer jacket of the cable is Santoprene
®
rubber.
This
compound
was
chosen
for
its
resistance
to
temperature extremes, moisture, and UV degradation.
However, this jacket will support combustion in air.
It is
rated as slow burning when tested according to U.L. 94
H.B. and will pass FMVSS302.
Local fire codes may
preclude its use inside buildings.
CAUTION