Campbell Scientific TE525MM TE525, TE525WS, and TE525MM Texas Electronics Rain - Page 19
TE525, TE525MM
View all Campbell Scientific TE525MM manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 19 highlights
TE525 Tipping Bucket Rain Gage (3) Place the can in the top funnel of the rain gage and pour 16 fluid ounces (1 pint) of water into the can. (A 16 oz. soft drink bottle filled to within 2.5 inches of the top may be used for a rough field calibration. An exact volume will allow for a more precise calibration). (4) If it takes less than 45 minutes for this water to run out, the hole in the can is too large. (5) The following number of tips should occur: TE525, TE525MM 100 ± 3 TE525WS 57 ± 2 (6) Adjusting screws are located on the bottom adjacent to the large center drain hole. Adjust both screws the same number of turns. Rotation clockwise increases the number of tips per 16 oz. of water; counter clockwise rotation decreases the number of tips per 16 oz. of water. One half turn of both screws causes a 2% to 3% change. (7) Check and re-level the rain gage lid. Factory Calibration: If factory calibration is required, contact Campbell Scientific to obtain an RMA (see Warranty and Assistance at front of manual). 13