Celestron Celestron Labs CB2000CF Compound Microscope Celestron Labs CB2000CF - Page 2

Page 2 highlights

MICROSCOPE OPERATION Before viewing specimens, please read these sections thoroughly regarding focusing, changing power (magnification), using the stage and adjusting illumination. VIEWING A SPECIMEN The images you see in your microscope will be upside down and reversed right to left. Your microscope includes prepared slides to help you get started. Your microscope is equipped with a mechanical stage (6) with a stage holder clamp (11) and directional knobs (15). 1. Use the clamp lever to open the clamping arm of the stage holder clamp (11). 2. C arefully place a prepared specimen slide (3 in x 1 in/76.2 mm x 25.4 mm size) inside the holder and close the clamping arm against the slide. 3. U se the stage movement knobs (15) to position the specimen over the opening in the stage (6). The large stage movement knob moves the X-axis (forward and backward), while the small stage movement knob moves the Y-axis (side to side). NOTE: To position the specimen directly under the objective lens, close the opening on the iris diaphragm until it is almost completely closed by moving the small lever. You should see a small beam of light projected on to the specimen slide. Now use the stage movement knobs to move the specimen directly inside the beam of light. You are now ready to focus and view the specimen. Use caution to avoid damaging the slide or object. When using higher powers while focusing, make sure the objective lens (5) does not hit the slide or specimen. 1. Start with the lowest power (4x objective lens and WF 10x eyepieces). Rotate the nosepiece (4) to change the objective lens (5) until the 4x objective lens is directly over the specimen. 2. Rotate each eyepiece diopter (2) clockwise so that they are all the way down. 3. A djust the distance between the eyepieces by sliding the eyepiece in or out horizontally. Grasp the knurled portion of each side of the plate to adjust the plate. 4. While looking through the eyepieces (1), gradually turn the coarse focus knob (14) until the specimen comes into view. Use the fine focus knob (16) to obtain a more precise focus. You may need to adjust the stage movement knobs (15) slightly to center the specimen in the field of view. 5. Adjust the eyepiece side plate until the whole field of view can be observed through both eyes at the same time without moving your head side to side. Depending on your individual eyes, you may need to make slight adjustments to the right and left eyepieces for the most comfortable viewing. Move the diopter (2) up or down until you have the specimen slide in sharp, comfortable focus. 6. F or higher powers, rotate the nosepiece (4) to change the objective lens (5) to 10x, 40x or 100x. This will yield a greater magnification. Gradually turn the coarse focus knob (14) to refocus on the specimen. NOTE: To be safe, you should first turn the coarse focus knob to lower the stage, before turning the objective carriage.) 7. You can replace the WF 10x eyepieces with the WF 20x eyepieces to obtain four additional high power magnifications, including (2000x), the highest power possible. ADJUSTING THE LIGHTING Specimens of different sizes, thicknesses, and colors require different levels of illumination. There are three ways to change the amount of illumination when viewing a specimen: adjusting the illumination using the illumination adjustment wheel (17), adjusting the Abbe condenser (8) and adjusting the iris diaphragm (7). ADJUSTING THE ILLUMINATION When viewing a specimen that is not transparent or dark in color you may need to increase the amount of light to resolve certain features or details. To do this, increase the brightness of the illuminator by turning the illuminator adjustment wheel (17) all the way to its highest setting. ADJUSTING THE CONDENSER When viewing with lower power (4x and 10x) objective lenses, you will need to lower the condenser lens in order to spread the light over the larger field of view. To change the position of the condenser (8), rotate the Abbe adjustment knob until the beam of light spreads wide enough to illuminate the entire field of view. ADJUSTING THE IRIS DIAPHRAGM As you lower the condenser (8) to spread out the light or change to a higher power objective lens, your image will appear dimmer. Instead of increasing the light intensity of the illuminator (which may "wash out" fine detail of the specimen you are viewing), open the aperture of the iris diaphragm (7) by moving the lever to let in more light. Opening and closing the diaphragm (7) will give a relief view of the specimen and allow you to change the depth of field of the specimen being viewed. USING FILTERS To bring out different levels of detail, experiment with changing the lighting color. To change colors, open the filter holder on the bottom of the diaphragm (7) by pushing it counterclockwise. Place the color filter in the filter holder and close it. You may need to refocus by adjusting the focus knobs. CARE, MAINTENANCE AND WARRANTY Your Celestron Labs microscope is a precision optical instrument and should be treated with care at all times. Follow these care and maintenance suggestions and your microscope will need very little maintenance throughout its lifetime. • When you are done using your microscope, remove any specimens left on the stage. • Turn off the power and unplug the power cord (if applicable). • Always place the plastic bag or dust cover over the microscope when not in use to help keep it clean. • Store the microscope in a dry, clean place. • Be very careful if using your microscope in direct sunlight to prevent damage to the microscope or your eyes. • To carry your microscope, place one hand on the "arm" of the microscope and the other hand under the base for support. Do not grasp your microscope by the focuser knob or head. • Clean the outside surfaces (metal and plastic) with a moist cloth. • Always unplug any cords before cleaning. • Never clean optical surfaces with cloth or paper towels as they can scratch optical surfaces easily. Instead, use an air blower or camel hair brush. • To clean fingerprints off of optical surfaces, use a lens cleaning agent and lens tissue available at most photo supply stores. When cleaning, do not rub in circles as this may cause streaks or scratches. • Never disassemble your microscope or clean internal optical surfaces. Only qualified technicians at the factory or an authorized repair facility should attempt these repairs. • When handling glass specimen slides, use care, as the edges can be sharp. YOUR MICROSCOPE HAS A TWO YEAR LIMITED WARRANTY. for more detailed information, please visit www.CelestronLabs.com Torrance, CA 90503 TEL (800) 421-9649 www.CelestronLabs.com Copyright 2014 Celestron All rights reserved. (Products or instructions may change without notice or obligation.) Designed and intended for those 14 years of age and older. Français Félicitations pour l'acquisition de votre microscope Celestron Labs. Votre microscope est un instrument optique de précision, fabriqué avec des matériaux de la plus grande qualité afin d'assurer sa durabilité et sa longue durée d'utilisation. Il est conçu pour vous offrir une vie entière de découvertes avec un niveau d'entretien minimal. Le microscope CB2000CF offre des puissances d'agrandissement de 40x à 2000x. Il convient parfaitement à l'observation d'échantillons de levures et de moisissures, de cultures, d'éléments végétaux et animaux, de fibres, de bactéries et autres. Avant de commencer à utiliser votre microscope Celestron Labs, lisez ces instructions pour vous familiariser avec les composants et les fonctions du microscope. Référez-vous au schémas du microscope pour repérer les pièces mentionnées dans le présent manuel. La dernière section du mode d'emploi donne des conseils d'entretien et de maintenance faciles à appliquer. DANS LA BOÎTE - Microscope CL-S20 - 4 lentilles d'objectifs: 4x, 10x, 40x (à ressort), 100x (à ressort) - 2 ensembles d'oculaires à angle large: (2) 10x avec pointeur , (2) 20x - 10 lamelles préparées - 3 filtres colorés

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8

MICROSCOPE OPERATION
Before viewing specimens, please read
these sections thoroughly regarding
focusing, changing power (magnification),
using the stage and adjusting illumination.
VIEWING A SPECIMEN
The images you see in your microscope will
be upside down and reversed right to left.
Your microscope includes prepared slides to
help you get started.
Your microscope is equipped with a
mechanical stage (6) with a stage holder
clamp (11) and directional knobs (15).
1.
Use the clamp lever to open the clamping
arm of the stage holder clamp (11).
2.
Carefully place a prepared specimen slide
(3 in x 1 in/76.2 mm x 25.4 mm size)
inside the holder and close the clamping
arm against the slide.
3.
Use the stage movement knobs (15) to
position the specimen over the opening in
the stage (6). The large stage movement
knob moves the X-axis (forward and
backward), while the small stage
movement knob moves the Y-axis (side
to side).
NOTE
:
To position the specimen directly under
the objective lens, close the opening
on the iris diaphragm until it is almost
completely closed by moving the small
lever. You should see a small beam
of light projected on to the specimen
slide. Now use the stage movement
knobs to move the specimen directly
inside the beam of light.
You are now ready to focus and view the
specimen. Use caution to avoid damaging
the slide or object. When using higher
powers while focusing, make sure the
objective lens (5) does not hit the slide or
specimen.
1.
Start with the lowest power (4x objective
lens and WF 10x eyepieces). Rotate the
nosepiece (4) to change the objective
lens (5) until the 4x objective lens is
directly over the specimen.
2.
Rotate each eyepiece diopter (2)
clockwise so that they are all the way
down.
3.
Adjust the distance between the
eyepieces by sliding the eyepiece in
or out horizontally. Grasp the knurled
portion of each side of the plate to adjust
the plate.
4.
While looking through the eyepieces (1),
gradually turn the coarse focus knob (14)
until the specimen comes into view. Use
the fine focus knob (16) to obtain a more
precise focus. You may need to adjust the
stage movement knobs (15) slightly to
center the specimen in the field of view.
5.
Adjust the eyepiece side plate until the
whole field of view can be observed
through both eyes at the same time
without moving your head side to side.
Depending on your individual eyes, you
may need to make slight adjustments
to the right and left eyepieces for the
most comfortable viewing. Move the
diopter (2) up or down until you have
the specimen slide in sharp, comfortable
focus.
6.
For higher powers, rotate the nosepiece
(4) to change the objective lens (5)
to 10x, 40x or 100x. This will yield a
greater magnification. Gradually turn the
coarse focus knob (14) to refocus on the
specimen.
NOTE
:
To be safe, you should first turn the
coarse focus knob to lower the stage,
before turning the objective carriage.)
7.
You can replace the WF 10x eyepieces
with the WF 20x eyepieces to obtain four
additional high power magnifications,
including (2000x), the highest power
possible.
ADJUSTING THE LIGHTING
Specimens of different sizes, thicknesses,
and colors require different levels of
illumination. There are three ways to change
the amount of illumination when viewing
a specimen: adjusting the illumination
using the illumination adjustment wheel
(17), adjusting the Abbe condenser (8) and
adjusting the iris diaphragm (7).
ADJUSTING THE
ILLUMINATION
When viewing a specimen that is not
transparent or dark in color you may need
to increase the amount of light to resolve
certain features or details. To do this,
increase the brightness of the illuminator
by turning the illuminator adjustment wheel
(17) all the way to its highest setting.
ADJUSTING THE
CONDENSER
When viewing with lower power (4x and
10x) objective lenses, you will need to lower
the condenser lens in order to spread the
light over the larger field of view. To change
the position of the condenser (8), rotate
the Abbe adjustment knob until the beam of
light spreads wide enough to illuminate the
entire field of view.
ADJUSTING THE IRIS
DIAPHRAGM
As you lower the condenser (8) to spread
out the light or change to a higher power
objective lens, your image will appear
dimmer. Instead of increasing the light
intensity of the illuminator (which may
“wash out” fine detail of the specimen
you are viewing), open the aperture of the
iris diaphragm (7) by moving the lever to
let in more light. Opening and closing the
diaphragm (7) will give a relief view of
the specimen and allow you to change the
depth of field of the specimen being viewed.
USING FILTERS
To bring out different levels of detail,
experiment with changing the lighting color.
To change colors, open the filter holder on
the bottom of the diaphragm (7) by pushing
it counterclockwise. Place the color filter in
the filter holder and close it. You may need
to refocus by adjusting the focus knobs.
CARE, MAINTENANCE AND
WARRANTY
Your Celestron Labs microscope is a
precision optical instrument and should be
treated with care at all times. Follow these
care and maintenance suggestions and your
microscope will need very little maintenance
throughout its lifetime.
• When you are done using your
microscope, remove any specimens left
on the stage.
• Turn off the power and unplug the power
cord (if applicable).
• Always place the plastic bag or dust cover
over the microscope when not in use to
help keep it clean.
• Store the microscope in a dry, clean place.
• Be very careful if using your microscope
in direct sunlight to prevent damage to
the microscope or your eyes.
• To carry your microscope, place one hand
on the “arm” of the microscope and the
other hand under the base for support. Do
not grasp your microscope by the focuser
knob or head.
• Clean the outside surfaces (metal and
plastic) with a moist cloth.
• Always unplug any cords before cleaning.
• Never clean optical surfaces with cloth or
paper towels as they can scratch optical
surfaces easily. Instead, use an air blower
or camel hair brush.
• To clean fingerprints off of optical
surfaces, use a lens cleaning agent
and lens tissue available at most photo
supply stores. When cleaning, do not rub
in circles as this may cause streaks or
scratches.
• Never disassemble your microscope or
clean internal optical surfaces. Only
qualified technicians at the factory or an
authorized repair facility should attempt
these repairs.
• When handling glass specimen slides, use
care, as the edges can be sharp.
YOUR MICROSCOPE HAS A
TWO YEAR LIMITED WARRANTY.
for more detailed information, please visit
www.CelestronLabs.com
Torrance, CA 90503
TEL
(800) 421-9649
www.CelestronLabs.com
Copyright 2014 Celestron
All rights reserved.
(Products or instructions may change
without notice or obligation.)
Designed and intended for those 14 years of
age and older.
Français
Félicitations pour l’acquisition de votre
microscope Celestron Labs. Votre
microscope est un instrument optique de
précision, fabriqué avec des matériaux de
la plus grande qualité afin d’assurer sa
durabilité et sa longue durée d’utilisation.
Il est conçu pour vous offrir une vie entière
de découvertes avec un niveau d’entretien
minimal.
Le microscope CB2000CF offre des
puissances d’agrandissement de 40x
à 2000x. Il convient parfaitement à
l’observation d’échantillons de levures et
de moisissures, de cultures, d’éléments
végétaux et animaux, de fibres, de bactéries
et autres.
Avant de commencer à utiliser votre
microscope Celestron Labs, lisez ces
instructions pour vous familiariser avec les
composants et les fonctions du microscope.
Référez-vous au schémas du microscope
pour repérer les pièces mentionnées dans
le présent manuel. La dernière section
du mode d’emploi donne des conseils
d’entretien et de maintenance faciles à
appliquer.
DANS LA BOÎTE
- Microscope CL-S20
- 4 lentilles d’objectifs: 4x, 10x, 40x (à
ressort), 100x (à ressort)
- 2 ensembles d’oculaires à angle large: (2)
10x avec pointeur , (2) 20x
- 10 lamelles préparées
- 3 filtres colorés