Fluke 123B/S Service Manual - Page 72
Input A and B DC Voltage Accuracy Test, WARNING
View all Fluke 123B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 72 highlights
123/124 Service Manual 20. Verify that no trace is shown on the test tool display, and that the status line at the display bottom shows Wait:B . If the display shows the traces and status Hold:B , then press to re-arm the test tool for a trigger. 21. Increase the 5500A voltage slowly in 0.1V steps, using the 5500A EDIT FIELD function, until the test tool is triggered, and the traces are shown. 22. Verify that the 5500A voltage is between +1.5V and +2.5V when the test tool is triggered. To repeat the test, start at step 19. 23. Set the 5500A to Standby. 24. Press to clear the display. 25. Press to enable the arrow keys for Trigger Level and Slope adjustment. 26. Using select negative slope triggering ( ). 27. Using set the trigger level to +2 divisions from the screen center. For negative slope triggering, the trigger level is the bottom of the trigger icon ( ). 28. Set the 5500A to source +3V DC. 29. Verify that no trace is shown on the test tool display, and that the status line at the display bottom shows Wait:B . If the display shows the traces and status Hold:B , then press to re-arm the test tool for a trigger. 30. Decrease the 5500A voltage in 0.1V steps, using the 5500A EDIT FIELD function, until the test tool is triggered, and the traces are shown. 31. Verify that the 5500A voltage is between +1.5V and +2.5V when the test tool is triggered. To repeat the test, start at step 28. 32. When you are finished, set the 5500A to Standby. 4.5.9 Input A and B DC Voltage Accuracy Test WARNING Dangerous voltages will be present on the calibration source and connecting cables during the following steps. Ensure that the calibrator is in standby mode before making any connection between the calibrator and the test tool. Proceed as follows: 1. Connect the test tool to the 5500A as for the previous test (see Figure 4-5). 2. Select the following test tool setup: • Press select auto ranging (AUTO in top of display). • Press to open the INPUT A MEASUREMENTS menu, and choose: MEASURE on A: I VDC • Press to open the INPUT B MEASUREMENTS menu, and choose: INPUT B: I ON | MEASURE on B: I VDC 4-14