Fluke 123B/S Service Manual - Page 82
Video Trigger Test, Test Tool Input A to TV Signal Generator
View all Fluke 123B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 82 highlights
123/124 Service Manual Table 4-9. Capacitance Measurement Verification Points 5500A output Reading 40 nF 300 nF 3 µF 30 µF 300 µF 0 (remove test tool input connections ) 39.10 to 40.90 293.0 to 307.0 2.930 to 3.070 29.30 to 30.70 293.0 to 307.0 00.00 to 00.10 (see steps 7...10) 4.5.19 Video Trigger Test Only one of the systems NTSC, PAL, or SECAM has to be verified. Proceed as follows: 1. Connect the test tool to the VIDEO output of the TV Signal Generator as shown in Figure 4-8. Figure 4-8. Test Tool Input A to TV Signal Generator 2. Select the following test tool setup: • Reset the test tool (power off and then on with ). • Press to open the SCOPE INPUTS menu. • Press to open the TRIGGER menu and choose: I VIDEO on A... From the shown VIDEO TRIGGER menu choose: SYSTEM: I NTSC or I PAL or I SECAM LINE: I SELECT POLARITY: I POSITIVE • Using set the Input A sensitivity to 200 mV/div. ST8141.WMF 4-24