Fluke 1520 FE 1520 Users Manual - Page 37

Analog Bar Graph, Short Circuit Current

Page 37 highlights

Fluke 1520 Users Manual Test Voltages Accuracy Nominal Current Number of measurements Circuitry Protection Display Range Measurement Range Accuracy Resolution Analog Bar Graph Open Circuit Voltage Short Circuit Current Test Leads Zero Number of measurements Circuitry Protection 32 250 V, 500 V, 1000 V +20 %, -0 % 1 mA 5,000 test inhibited if ≥ 30 V ac or dc at inputs Lo Ω 40.00 Ω 0.10 Ω to 40.00 Ω 2 % + 2 counts 0.01 Ω 0 to 100 Ω 6 V typical 200 mA minimum, (0.2 to 2.0 Ω for >20% battery capacity remaining). 2 Ω 5,000 test inhibited if ≥ 30 V ac or dc at inputs