Fluke 199C/S FE 192,196,199 C Users Manual - Page 117
Probe, Safety, Electrical specifications, Accuracy, Environmental
View all Fluke 199C/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 117 highlights
10:1 Probe Safety Max. Input Voltage 1000 V CAT II 600 V CAT III Max. Floating Voltage from any terminal to earth ground 1000 V CAT II 600 V CAT III Electrical specifications Input Impedance at probe tip ......10 MΩ ±2 %//14 pF ±2 pF Capacity Adjustment Range 10 to 22 pF Attenuation at DC (1 MΩ input 10 x Bandwidth (with FLUKE 199C) .....DC to 200 MHz (-3 dB) Accuracy Probe accuracy when adjusted on the test tool: DC to 20 kHz 1 % 20 kHz to 1 MHz 2 % 1 MHz to 25 MHz 3 % For higher frequencies the probe's roll off starts affecting the accuracy 9 Specifications 10:1 Probe Environmental Temperature Operating 0 to 50 °C (32 to 122 °F) Storage 20 to +60 °C (-4 to +140 °F) Altitude Operating 3 km (10 000 feet) Storage 12 km (40 000 feet) Humidity Operating at 10 to 30 °C (50 to 86 °F 95 % MAX. INPUT VOLTAGE (Vrms) 1000 500 CAT II CAT III 200 100 50 20 10 5 2 1 0.01 0.02 0.05 0.1 0.2 0.5 1 2 5 10 20 50 100 200 FREQUENCY (MHz) Figure 51. Max. Voltage From Probe Tip to earth ground and From Probe Tip to Probe Reference 109