Fluke 199C/S FE 192,196,199 C Users Manual - Page 118
Electromagnetic Immunity, Meter Mode Vdc, Vac, Vac+dc
View all Fluke 199C/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 118 highlights
Fluke 192B - 196B/C - 199B/C Users Manual Electromagnetic Immunity The Fluke 190 series, including standard accessories, conforms with the EEC directive 89/336 for EMC immunity, as defined by EN-61326-1, with the addition of the following tables. Scope Mode (10 ms/div): Trace disturbance with VPS200 voltage probe shorted Table 1 No visible disturbance Frequency range 10 kHz to 20 MHz Frequency range 20 MHz to 100 MHz Frequency range 100 MHz to 1 GHz E = 3V/m 2 mV/div to 100 V/div 200 mV/div to 100 V/div 500 mV/div to 100 V/div *) (*) With the 20 MHz Bandwidth Filter switched on: no visible disturbance. With the 20 MHz Bandwidth Filter switched off: disturbance is max 2 div. Table 2 Disturbance less than 10% of full scale Frequency range 20 MHz to 100 MHz E = 3V/m 10 mV/div to 100 mV/div Test Tool ranges not specified in tables 1 and 2 may have a disturbance of more than 10% of full scale. Meter Mode (Vdc, Vac, Vac+dc, Ohm and Continuity): Reading disturbance with test leads shorted Table 3 Disturbance less than 1% of full scale Frequency range 10 kHz to 1 GHz E = 3V/m 500mV to 1000V , 500Ohm to 30 MOhm ranges 110