AC Current (Flexible Current Probe) - parts
View all Fluke 381 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 26 highlights
381 Users Manual AC Current (Flexible Current Probe) XW Warning To prevent possible electrical shock or personal injury: Do not apply the Flexible Current Probe around or remove from HAZARDOUS LIVE conductors. Take special care during fitting and removal of the Flexible Current Probe. De-energize the installation under test or wear suitable protective clothing. To use the Flexible Current Probe, follow these instructions: 1. Connect the Flexible Current Probe to the Meter. See Figure 5. 2. Connect the flexible part of the Flexible Current Probe around the conductor. If opening the end of the Flexible Current Probe to make the connection, make sure that you close and latch it. See the detail in Figure 5. You should be able to hear and feel the Flexible Current Probe lock snap into place. Note When measuring current, center the conductor in the Flexible Current Probe. If possible, avoid taking measurements close to other current-carrying conductors. 3. Keep the probe coupling more than 2.5 cm (1 inch) away from the conductor. D 4. Turn the Rotary Function Switch to . When the Rotary Function Switch is in the correct position, Y shows on the display, meaning that the readings are coming from the Flexible Current Probe. 22