Initial Conditions and Data Entries, Starting and Stopping Logging
View all Fluke 54-II-B manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 20 highlights
53IIB/54IIB Users Manual Initial Conditions and Data Entries Logged readings include initial conditions and data entries. The initial conditions are the thermocouple type and the offsets for each thermocouple input. You can only view initial conditions using FlukeView Forms. The data entries are a time stamp, the T1 reading, and the T2 and T1-T2 readings (Model 54). You can view these values by pressing or using FlukeView Forms. Temperature readings display 0.1 degree resolution in FlukeView Forms. The thermometer has 500 memory locations. The thermometer stores 499 sets of temperature readings and one set of initial conditions when logging continuously. It stores 250 sets of temperature readings and 250 sets of initial conditions when logging individual points manually. Starting and Stopping Logging Setup, memory clear, and PC communications are inaccessible during logging. 1. Set the logging interval. (See "Changing Setup Options.") 2. Press to start logging. The display shows . 3. Press again to stop logging. 4. If you selected a manual logging interval, press each time you want to store logged readings in memory. 16 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664