574 Warning for the Model 574 NI Concerning Factory Mutual Approved Nonincendive Devices: Operation in Environments that Require Nonincendive Devices WARNING IN HAZARDOUS LOCATIONS DO NOT use serial port connections, change batteries or open serial port cover. To reduce risk of explosion in hazardous locations, use only Fluke temp probe part 2432508 and do not use other accessories, such as power supply and cables. A nonincendive rating (NI) indicates that this infrared thermometer has been tested to standards for preventing explosions in hazardous areas by limiting the ability of equipment to ignite a specified flammable gas or vapor-in-air mixture. Nonincendive equipment is incapable of releasing sufficient electrical or thermal energy to ignite flammable gases or vapors under NORMAL operation and environmental conditions. This noncontact thermometer has a Factory Mutual Nonincendive rating. The rating from this USA organization reads: "Nonincendive, Class I, Division 2, Groups A, B, C, D; Class I, Zone 2 IIC; T4 Ta = 50ºC when used with 1.5V alkaline batteries." A Class I, Division 2 location is a location: • where volatile flammable liquids or flammable gases or vapors exist, but are normally confined within closed containers; • where ignitable concentrations of gases, vapors or liquids are normally prevented by positive mechanical ventilation; or • adjacent to a Class I, Division 1 location, where ignitable concentrations might be occasionally communicated • groups A, B, C, D refers to: Acetylene, Hydrogen, Ethylene, and Propane. 6

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32

6
574
Concerning Factory Mutual Approved
Nonincendive Devices:
Operation in Environments that Require
Nonincendive Devices
WARNING
IN HAZARDOUS LOCATIONS DO NOT use serial port con-
nections, change batteries or open serial port cover.
To re-
duce risk of explosion in hazardous locations, use only Fluke
temp probe part 2432508 and do not use other accessories,
such as power supply and cables.
A nonincendive rating (NI) indicates that this infrared thermo-
meter has been tested to standards for preventing explosi-
ons in hazardous areas by limiting the ability of equipment
to ignite a specified flammable gas or vapor-in-air mixture.
Nonincendive equipment is incapable of releasing sufficient
electrical or thermal energy to ignite flammable gases or
vapors under NORMAL operation and environmental condi-
tions.
This noncontact thermometer has a Factory Mutual Nonin-
cendive rating.
The rating from this USA organization reads:
“Nonincendive, Class I, Division 2, Groups A, B, C, D; Class
I, Zone 2 IIC; T4 Ta = 50ºC when used with 1.5V alkaline
batteries.“
A Class I, Division 2 location is a location:
• where volatile flammable liquids or flammable gases
or vapors exist, but are normally confined within
closed containers;
• where ignitable concentrations of gases, vapors or liquids
are normally prevented by positive mechanical
ventilation; or
• adjacent to a Class I, Division 1 location, where ignitable
concentrations might be occasionally communicated
• groups A, B, C, D refers to: Acetylene, Hydrogen,
Ethylene, and Propane.
Warning for the Model 574 NI