Write LRV and URV Values, Write LRV, Write LRV and URV Screen, Function Select, AUTOMATIC
View all Fluke 710 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 30 highlights
Shop for Fluke products online at: www. Write LRV and URV Values Note For the screens in this section, tag names can be truncated to fit on one line. flashes in the upperright corner to show a live connection. If the HART write commands are not active, these functions are not available and an error message shows instead of the screen in Figure 32. Precision Loop Calibrator Write LRV and URV Values Write LRV The Product warns you to change the loop to MANUAL before you proceed. Push the selection knob to proceed. Push to go to the Write LRV and Write URV menu. The present LRV value and units show. See Figure 33. Figure 32. Write LRV and URV Screen Use the selection knob to choose the function. Push the selection knob to choose the highlighted action and go to the corresponding screen. See Write LRV and Write URV. Push to go to the Function Select menu. Figure 33. Write LRV Screen 1. Push the selection knob to move the decade cursor. 2. Turn the selection knob to increment or decrement the value in steps indicated by the selected decade. 3. Push and hold the knob to send the new value to the HART device. An error shows if the HART device rejects the value. 4. Push to go to the Function Select menu. A reminder to change the loop to AUTOMATIC shows first. MyFlukeStore.com 1.888.610.7664 25