Fluke 710 Product Manual - Page 33
Trim, Set, and Zero Menu, Trim 4 mA, Trim, and Zero Screen, Set and Zero
View all Fluke 710 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 33 highlights
Shop for Fluke products online at: www. 709/709H/710 Users Manual Trim, Set, and Zero Menu Note For the screens in this section, tag names can be truncated to fit on one line. flashes in the upper-right corner to show a live connection. If the HART write commands are not enabled, these functions are not available and an error message shows instead of the screen shown in Figure 38. Trim 4 mA If the operation mode is Communicator Only, this function is not available and an error message shows. A prompt to change the loop to MANUAL before you proceed shows. Push the selection knob to proceed. Push to go to the Trim, Set and Zero menu. While the HART device is changed to fixed output mode, an error shows if the HART device rejects the mode change command. When the mode change is successful, the screen in Figure 39 shows. MyFlukeStore.com 1.888.610.7664 Figure 38. Trim, Set, and Zero Screen Use the selection knob to select and start the action. Push to go to the Function Select menu. Figure 39. Trim 4 mA Screen While the output to settles at 4 mA, the screen shows the Product measurement. The measurement updates once a second. 1. Push the selection knob to trim the HART device. Stay on this screen to evaluate the result. An error shows if the HART device rejects the trim command. 2. Push to change the HART device to normal output mode and go to the Trim, Set, and Zero menu. A warning to change the loop to AUTOMATIC shows first. An error shows if the device rejects the mode change command. 28