Fluke 87V-MAX Product Manual - Page 14
Autoranges to the 400 mV, 4V, 40V, 400V or 1000V, Table 3. Beeper Response in Continuity Test
View all Fluke 87V-MAX manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 14 highlights
OFF Power to the meter is turned off. K Volts ac Autoranges to the 400 mV, 4V, 40V, 400V or 1000V range. L Volts dc Autoranges to the 4V, 40V, 400V or 1000V range. l Millivolts dc Single 400 mV range. ReE Resistance (Ω), conductance (1/Ω), capacitance or continuity R testing. Press BLUE button to toggle between the resistance and capacitance function. (The response of the display and the pushbuttons slows down in the capacitance mode.) Autoranges to the 400Ω, 4 kΩ, 40 kΩ, 400 kΩ, 4 MΩ, or 40 MΩ resistance range. How to Use the Meter In Manual Ranging mode, 40 nS conductance range (equal to a 25-100,000 MΩ range) is selectable. (See item 9.) Autoranges to the 05.00 nF, .0500 µF, 0.500 µF, or 05.00 µF capacitance range. When testing continuity, the beeper sounds if the resistance falls below the typical values indicated in Table 3. Table 3. Beeper Response in Continuity Test Input Range Beeper On If 400.0Ω 4.000 kΩ 40.00 kΩ 400.0 kΩ 4.000 MΩ 40.00 MΩ < 40Ω < 200Ω < 2 kΩ < 20 kΩ < 200 kΩ < 200 kΩ L Diode Test Measures forward voltage of semiconductor junction(s) at approximately 0.5 mA test current. Single 0-3V range. 7 MyFlukeStore Find QuaSlhitoypPforordFulucktespOrondliuncetsato:nline awt:ww.wGwlo. balTestSupply.com.com [email protected]