3 3.1 3.2 3 Specifications and Environmental Conditions Specifications Specifications and Environmental Conditions Specifications Operating Range Accuracy Stability Well-to-Well Uniformity Resolution Display Scale Thermal Electric Devices (TED) Heating Time Cooling Time Stabilization Time Immersion Depth Power Requirements Inserts Computer Interface Size Weight -25°C to 140°C (-13°F to 284°F), at 23°C ambient ±0.25°C (±1.0°C in holes > ¼" [6.4 mm]) ±0.02°C at -25°C ±0.04°C at 140°C ±0.1 with similar probes 0.1°C or °F °F or °C, switchable 150 W 18 minutes from ambient to 140°C 20 minutes from ambient to -25°C 7 minutes 4.875 inches (124 mm) 115 VAC (±10%), 1.3 A or 230 VAC (±10%) 0.7 A, switchable, 50/60 Hz, 150W A, B, or C RS-232 included with free Interface-it Software (Model 9930) 10.25" H x 5.63" W x 9.63" D (26.1 x 14.3 x 24.5 cm) 12 lb. (5.7 kg) Environmental Conditions Although the instrument has been designed for optimum durability and trouble-free operation, it must be handled with care. The instrument should not be operated in an excessively dusty or dirty environment. Maintenance and cleaning recommendations can be found in the Maintenance Section of this manual. The instrument operates safely under the following conditions: • temperature range: 5-45°C (41-113°F) • ambient relative humidity: 15 - 50% • pressure: 75kPa - 106kPa • mains voltage within ± 10% of nominal • vibrations in the calibration environment should be minimized • altitude less than 2,000 meters 9 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54

3
Specifications and Environmental
Conditions
3.1
Specifications
Operating Range
–25°C to 140°C (–13°F to 284°F), at 23°C ambient
Accuracy
±0.25°C (±1.0°C in holes > ¼” [6.4 mm])
Stability
±0.02°C at –25°C
±0.04°C at 140°C
Well-to-Well Uniformity
±0.1 with similar probes
Resolution
0.1°C or °F
Display Scale
°F or °C, switchable
Thermal Electric Devices (TED)
150 W
Heating Time
18 minutes from ambient to 140°C
Cooling Time
20 minutes from ambient to –25°C
Stabilization Time
7 minutes
Immersion Depth
4.875 inches (124 mm)
Power Requirements
115 VAC (±10%), 1.3 A or 230 VAC (±10%) 0.7 A, switchable, 50/60
Hz, 150W
Inserts
A, B, or C
Computer Interface
RS-232 included with free Interface
-it
Software (Model 9930)
Size
10.25” H x 5.63” W x 9.63” D
(26.1 x 14.3 x 24.5 cm)
Weight
12 lb. (5.7 kg)
3.2
Environmental Conditions
Although the instrument has been designed for optimum durability and trou-
ble-free operation, it must be handled with care. The instrument should not be
operated in an excessively dusty or dirty environment. Maintenance and clean-
ing recommendations can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
temperature range: 5–45°C (41–113°F)
ambient relative humidity: 15 - 50%
pressure: 75kPa - 106kPa
mains voltage within ± 10% of nominal
vibrations in the calibration environment should be minimized
altitude less than 2,000 meters
9