Setting the Temperature
View all Fluke 9103-A-156 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 20 highlights
9103 Dry-Well User's Guide 4.4 Setting the Temperature Section 7.2 explains in detail how to set the temperature set-point on the calibrator using the front panel keys. The procedure is summarized here. (1) Press "SET" twice to access the set-point value. (2) Press "UP" or "DOWN" to change the set-point value. (3) Press "SET" to program in the new set-point. (4) Press "EXIT" to return to the temperature display. When the set-point temperature is changed the controller switches the well TEDs on or off to raise or lower the temperature. The displayed well temperature gradually changes until it reaches the set-point temperature. The well may require 25 minutes to reach the set-point depending on the span. Another 6 to 10 minutes is required to stabilize within ±0.1°C of the set-point. Ultimate stability may take 20 to 30 minutes more of stabilization time. 12 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664