Fluke 9103-A-156 Product Manual - Page 20

Setting the Temperature

Page 20 highlights

9103 Dry-Well User's Guide 4.4 Setting the Temperature Section 7.2 explains in detail how to set the temperature set-point on the calibrator using the front panel keys. The procedure is summarized here. (1) Press "SET" twice to access the set-point value. (2) Press "UP" or "DOWN" to change the set-point value. (3) Press "SET" to program in the new set-point. (4) Press "EXIT" to return to the temperature display. When the set-point temperature is changed the controller switches the well TEDs on or off to raise or lower the temperature. The displayed well temperature gradually changes until it reaches the set-point temperature. The well may require 25 minutes to reach the set-point depending on the span. Another 6 to 10 minutes is required to stabilize within ±0.1°C of the set-point. Ultimate stability may take 20 to 30 minutes more of stabilization time. 12 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54

4.4
Setting the Temperature
Section 7.2 explains in detail how to set the temperature set-point on the cali-
brator using the front panel keys. The procedure is summarized here.
(1) Press “SET”
twice
to access the set-point value.
(2) Press “UP” or “DOWN” to change the set-point value.
(3) Press “SET” to program in the new set-point.
(4) Press “EXIT” to return to the temperature display.
When the set-point temperature is changed the controller switches the well
TEDs on or off to raise or lower the temperature. The displayed well tempera-
ture gradually changes until it reaches the set-point temperature. The well may
require 25 minutes to reach the set-point depending on the span. Another 6 to
10 minutes is required to stabilize within ±0.1°C of the set-point. Ultimate sta-
bility may take 20 to 30 minutes more of stabilization time.
9103 Dry-Well
User’s Guide
12