HoMedics BPA-100 User Manual - Page 16

Specifications, Troubleshooting

Page 16 highlights

TROUBLESHOOTING If any abnormality arises during use, please check the following points SYMPTOMS POSSIBLE CAUSES Unit does not turn on when Batteries have the START/STOP run down. button is pushed. Battery polarities have been positioned incorrectly. CORRECTION Replace them with four new AA alkaline batteries. Re-insert the batteries in the correct positions. EE measurement error Cuff has been placed symbol shown on display or incorrectly. the blood pressure value is displayed excessively low (or high). Did you talk or move during measurement? Shaking of the arm with the cuff on. Wrap the cuff properly so that it is positioned correctly. Measure again. Keep arm steady during measurement. Er error symbol shown on display - measurement result exceeds measurement range. Did you move during meas- Measure again. Do not urement? move during measurement. Shaking of the arm with the cuff on. E1 error symbol shown on Air circuit abnormality. Cuff Check cuff connection. display tube may not be plugged Measure again. into monitor correctly. E2 error symbol shown on Inflation pressure exceeding Switch the unit off, then display 300 mmHg. measure again. E3 error symbol shown on Data error. display Remove and reload the batteries. - - appears on display Recalibration alert. Refer to page 25 for more information. Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no circumstance should you disassemble or attempt to repair the unit by yourself. Contact information for HoMedics Consumer Relations Department can be found on the warranty page. 29 SPECIFICATIONS Power Source: Four 1.5V DC (AA) batteries or optional 12V DC AC adapter (not included) Measurement Method: Oscillometric Measurement Range: Pressure: 40~250 millimeters mercury (mmHg) Pulse: 40~199 beats/minute Accuracy: Pressure: ±3 mmHg Pulse: ±5% of reading Pressure Sensor: Inflation: Deflation: Memory Capacity: Auto-shut-off: Operation Environment: Storage Environment: Weight: Semi-conductor Pump driven Automatic Pressure release valve 30 memories for each User (60 total) 1 minute after last button operation Temperature: 50°F~104°F (10°C ~40°C) Humidity: 40~85% RH max. Temperature: 14°F~140°F (-10°C~60°C) Humidity: 10~90% RH max. 0.65 lbs (without batteries) Arm Circumference: Dimensions: 8.7"~18.1" (22~46cm) 4.3"(L) x 6"(W) x 2.9"(H) 109mm(L) x 153mm(W) x 74 mm(H) Type BF Device and cuff are designed to provide special protection against electrical shocks. Note: These specifications are subject to change without notice. 30

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32

30
Power Source:
Four 1.5V DC (AA) batteries or
optional 12V DC AC adapter (not included)
Measurement Method:
Oscillometric
Measurement Range:
Pressure: 40~250 millimeters mercury (mmHg)
Pulse: 40~199 beats/minute
Accuracy:
Pressure: ±3 mmHg
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Pump driven
Deflation:
Automatic Pressure release valve
Memory Capacity:
30 memories for each User (60 total)
Auto-shut-off:
1 minute after last button operation
Operation Environment:
Temperature:
50°F~104°F (10°C ~40°C)
Humidity:
40~85% RH max.
Storage Environment:
Temperature:
14°F~140°F (-10°C~60°C)
Humidity:
10~90% RH max.
Weight:
0.65 lbs (without batteries)
Arm Circumference
:
8.7"~18.1" (22~46cm)
Dimensions:
4.3"(L) x 6"(W) x 2.9"(H)
109mm(L) x 153mm(W) x 74 mm(H)
Type BF
Device and cuff are designed to provide special
protection against electrical shocks.
SPECIFICATIONS
Note:
These specifications are subject to change without notice.
29
TROUBLESHOOTING
If any abnormality arises during use, please check the following points
Note:
If the unit still does not work, contact HoMedics Consumer
Relations.
Under no circumstance should you disassemble or
attempt to repair the unit by yourself.
Contact information for
HoMedics Consumer Relations Department can be found on
the warranty page.
SYMPTOMS
POSSIBLE
CAUSES
CORRECTION
Unit does not turn on when
the START/STOP
button is pushed.
Batteries have
run down.
Replace them with
four new AA
alkaline batteries.
Battery polarities
have been positioned
incorrectly.
Re-insert the batteries in the
correct positions.
EE measurement error
symbol shown on display or
the blood pressure value is
displayed excessively low
(or high).
Cuff has been placed
incorrectly.
Wrap the cuff properly
so that it is positioned
correctly.
Did you talk or move
during measurement?
Measure again.
Keep arm
steady during
measurement.
Shaking of the arm with
the cuff on.
Er error symbol shown on
display - measurement
result exceeds measure-
ment range.
Did you move during meas-
urement?
Shaking of the arm with
the cuff on.
Measure again. Do not
move during measurement.
E1 error symbol shown on
display
Air circuit abnormality.
Cuff
tube may not be plugged
into monitor correctly.
Check cuff connection.
Measure again.
E2 error symbol shown on
display
Inflation pressure exceeding
300 mmHg.
Switch the unit off, then
measure again.
E3 error symbol shown on
display
Data error.
Remove and reload the
batteries.
– – appears on display
Recalibration alert.
Refer to page 25 for more
information.