Invacare M41RSOLID20R Owners Manual - Page 59
Removing/Installing the Batteries
View all Invacare M41RSOLID20R manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 59 highlights
SECTION 13-BATTERIES ƽ WARNING Batteries with terminal configuration as shown below MUST be used. Batteries that have the reverse terminal configuration MUST NOT be used - otherwise injury and damage may occur. USE THIS CONFIGURATION U1 Battery NEGATIVE (-) Battery Terminal U1 Battery DO NOT USE POSITIVE (+) Battery Terminal NEGATIVE (-) Battery Terminal POSITIVE (+) Battery Terminal NOTE: Recommended battery type is spill proof and requires no maintenance except routine charging. NOTE: Charge batteries daily. DO NOT allow batteries to completely discharge. To charge batteries, refer to Charging Batteries on page 63. NOTE: Invacare recommends that both batteries be replaced if one battery is defective. Removing/Installing the Batteries NOTE: For this procedure, refer to FIGURE 13.1 on page 60. ƽ WARNING Install protective caps on POSITIVE (+) and NEGATIVE (-) terminals. All battery terminal covers (two on the right battery and two on the left battery) MUST be installed prior to use. Removing 1. Disconnect the joystick. Refer to Disconnecting/Connecting the Joystick on page 57. 2. Remove the seat. Refer to Removing/Installing the Seat Assembly on page 38. 3. Remove the top shroud. Refer to Removing/Installing the Top Shroud on page 50. 4. Disconnect the right battery from the controller (BLACK connector). Part No. 1143206 59 Pronto® M41 with SureStep®