Accuracy Check Vertical plumb points Checking the accuracy of the upper plumb point: Checking the accuracy of the lower plumb point: Place the lock switch in the "Unlocked" position (see Levelling lock). Set up the laser on its tripod to wall mount bracket near point A1 at a minimum distance of 1.5 m from point B1. The horizontal laser is aligned in direction 1. Mark the laser points A1 and B1 with a pin. Rotate the instrument by 180° so that it points in the opposite direction 2 to direction 1. Adjust the instrument so that the laser beam hits point A1 exactly. If point B2 is no further than 2 mm away from point B1, then the instrument is within tolerance. Should your instrument be outside of the specified tolerance, please contact a local dealer or an authorised Makita distributor. Makita SK106D/SK106GD 16

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24

Accuracy Check
Vertical plumb points
Checking the accuracy of the upper plumb
point:
Place the lock switch in the "Unlocked" position
(see
Levelling lock
). Set up the laser on its tripod to
wall mount bracket near point A1 at a minimum dis-
tance of 1.5 m from point B1. The horizontal laser
is aligned in direction 1. Mark the laser points A1
and B1 with a pin.
Checking the accuracy of the lower plumb
point:
Rotate the instrument by 180° so that it points in
the opposite direction 2 to direction 1. Adjust the
instrument so that the laser beam hits point A1
exactly. If point B2 is no further than 2 mm away
from point B1, then the instrument is within tol-
erance.
Should your instrument be outside of the spe-
cified tolerance, please contact a local dealer
or an authorised Makita distributor.
Makita SK106D/SK106GD