Changing the Logging Interval, If you selected a user-defined logging interval - thermometer
View all Fluke 54 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 14 highlights
53 & 54 Series II Users Manual Changing the Logging Interval The logging interval determines how often the thermometer stores logged readings in memory. You choose the length of the logging interval. See "Using Memory." The thermometer stores logged readings at the end of each logging interval. You can select a logging interval of 1 second (1), 10 seconds (2), 1 minute (3), 10 minutes (4), or user-defined (USef). You can also set the logging interval manually (0). Each time you press J, the thermometer stores the current readings in memory. 1. Press K or N until the display shows INTERVAL. 2. Press E to display the logging interval choices. 3. Press K or N until the display shows the logging interval you want, and then press E to select. 4. If you selected a user-defined logging interval: • Press K or N until the display shows hour:min or min:sec, and then press E to select. The left two digits blink. • Press K or N until the left two digits you want appear on the display, and then press E to select them. The right two digits blink. • Press K or N until the right two digits you want appear on the display, and then press E to select them. Holding down K or N causes the number to change more quickly. 10