Fluke 54 FE 53 & 54 II Users Manual - Page 25
Replacement Parts and Accessories, Electrical cont. - ii service manual
View all Fluke 54 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 25 highlights
Electrical (cont.) Measurement Accuracy, T1, T2 or T1-T2 (Model J[rbe-,eaKldoi-wn, gT−-+1, 00E0.-3,oaCoCn(d−(01N.45-t8yopFoFe)])::±a[0d.d050.%15o%f 54) of reading for J-, K-, E-, and N-type; and 0.45 % of reading for T-type] R- and S-type: 0.4 oC (0.7 oF)] ±[0.05 % of reading + Temperature Coefficient (00.0.015%oFopf ereraodFi)nogu+ts0id.0e3thoeC per oC specified +18 oC to 28 oC (+64 oF to +82 oF) range [below −100 oC (−148 oF): add 0.04 % of reading for J-, K-, E-, and N-type; and 0.08 % of reading for T- type] Electromagnetic Susceptibility: ±2 oC (±3.6 oF) for 80 Compatibility MHz to 200 MHz in 1.5 V/m field, for 200 MHz to 1000 MHz in 3 V/m field. Emmisions: Commercial Limits per EN50081-1 Maximum Differential Common Mode Voltage 1 V (Maximum voltage difference between T1 and T2) Temperature Scale ITS-90 Applicable Standards NIST-175 Accuracy is specified for ambient temperatures between 18 oC (64 oF) and 28 oC (82 oF) for a period of 1 year. The above specifications do not include thermocouple error. 53 & 54 Series II Replacement Parts and Accessories Replacement Parts and Accessories Accessory Part Number Holster and Flex Stand™ Assembly 1272438 AA NEDA 15A IEC LR6 batteries 376756 80PK-1 K-Type Bead Thermocouple 773135 CD-ROM 1276106 Service Manual 1276123 21