Invacare MA65 Owners Manual - Page 26
Preparing for CPR Procedure, About Power Outage and Transportation
View all Invacare MA65 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 26 highlights
microAIR® MA65 Series 5.6 Preparing for CPR Procedure 1. Press and hold the tabs A on hose connector B while pulling the hose C from the control unit D. 2. Twist the CPR connector E at the foot end of the overlay F and all air will be evacuated in the torso and chest region of the overlay within 18 seconds. DANGER! Risk of Death or Injury To reduce the risk of death or injury: - Move patient to appropriate surface to perform CPR if unable to deflate mattress. 5.7 About Power Outage and Transportation If the hoses remain connected to the control unit, the mattress retains air during a power outage, during transportation, when the control unit is unplugged or when the control unit is turned Off. 26 1148137-H
microAIR® MA65 Series
5.6
Preparing for CPR Procedure
1.
Press and hold the tabs
A
on hose connector
B
while pulling the hose
C
from the control unit
D
.
2.
Twist the CPR connector
E
at the foot end of the overlay
F
and all air will be evacuated in the torso and chest region
of the overlay within 18 seconds.
DANGER!
Risk of Death or Injury
To reduce the risk of death or injury:
– Move patient to appropriate surface to perform CPR if unable to deflate mattress.
5.7
About Power Outage and Transportation
If the hoses remain connected to the control unit, the mattress retains air during a power outage, during transportation, when
the control unit is unplugged or when the control unit is turned Off.
26
1148137-H