SECTION 7-MAINTENANCE Actuator Cap TOP Bearing Washer Nylon Washers Flat Washer Cap Nut Washer Bearing Lift Arm Actuator Mounting Bracket Cap Flat Bushing Nylon Washer Screw Actuator Lead Routing Hole Washers Cap Bearing Bushing Nut Washer Cap Washer BOTTOM Bearing Screw Mast Assembly Actuator Mounting Bracket Electric Actuator FIGURE 7.1 Replacing the Electric Actuator Replacing the Leg Actuators NOTE: For this procedure, refer to FIGURE 7.2 on page 25. NOTE: If the legs of the patient lift do not open and close properly, it may be necessary to replace one or both of the leg actuators. 1. Test the operation of the legs: • Press the CLOSE button on the hand control to close the legs. • Press the OPEN button on the hand control to open the legs. 2. While holding on to the mast handles, carefully tip the patient lift back until it rests on the floor. NOTE: It is necessary to remove both small base covers before removing the large base cover. 3. Remove the sixteen screws that attach the small base covers to the base. 4. Remove the eight pan head screws that attach the large base cover to the base. NOTE: If the left or the right leg of the patient lift does not operate properly, perform the following steps on the side of the base necessary to replace the actuator. If both legs do not operate properly, perform this replacement procedure on both sides of the base. 5. Locate and disconnect the leg actuator lead wires. Roze™ Stand Up Patient Lift 24 Part No. 1150703