Campbell Scientific TGA200A TGA Series Trace Gas Analzyers - Page 145

TGA100 and TGA100A, Multimode, power, Multimode power, Settings > Laser > Other, Laser Settings

Page 145 highlights

Appendix C. Optimizing Laser Parameters NOTE NOTE other frequencies will generally not be absorbed. The measured transmittance at the center of the absorption line gives an estimate of the laser multimode power. Two different methods are used depending on the TGA model TGA100 and TGA100A TGA100s and TGA100As have a long sample cell and a short reference cell. The laser's multimode power can be estimated by temporarily putting reference gas in the long sample cell. This increases the path length by a factor of almost 34. This test is best performed in conditions in which there is not an excessive amount of absorption as this can lead to two possible problems. First, the absorption lines become broader, and absorption in the tails of the absorption line can reduce the response at the edges of the spectral scan that are assumed to be 100% transmittance. This will give an error in the estimate of multimode power, especially if there is another absorption line nearby. Second, too much absorption will increase the chances of absorbing the multimode power in some other absorption lines of the gas. The correct absorption can usually be achieved by adjusting the pressure in the sample cell. To begin, go to the Settings > Laser > Other window and set the Multimode power % to zero. Then note the reference transmittance in the Laser Settings window. The transmittance measurement is affected by detector nonlinearity (see Appendix D.4, Detector Linearity Coefficients). The reference detector is more linear because it has a smaller signal. Therefore this measurement should be based on the reference detector, not the sample detector. To achieve the optimum amount of absorption, first note the reference transmittance with reference gas in the reference cell only (the normal configuration). Normally, the reference gas concentration is chosen to give approximately 50% absorption. For this test, however, it is best to have between 70% and 80% absorption. Usually this can be accomplished by reducing the sample flow to reduce the pressure in the analyzer. When the proper absorption is achieved, record the sample cell pressure. It is acceptable for the transmittance to be less than 70%, but it must NOT be greater than 80%. To configure the instrument for the multimode test, turn off the sample pump, swap the reference gas and sample connections, and turn the pump back on. This will put the reference gas in the long sample cell and the air sample in the short reference cell. The total flow to the sample pump should be the same as before, so the pressure should also be the same. Verify the pressure is within 10% of the previous pressure. Adjust the laser modulation current as needed - the increased absorption may make the line wider. Record the reference transmittance, shown in the Laser Settings window. This is the estimate of the laser's multimode power. Enter this value as the Multimode power (%) parameter in the Settings > Laser > Other window. Verify the reference transmittance is near zero. C-13

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135
  • 136
  • 137
  • 138
  • 139
  • 140
  • 141
  • 142
  • 143
  • 144
  • 145
  • 146
  • 147
  • 148
  • 149
  • 150
  • 151
  • 152
  • 153
  • 154
  • 155
  • 156
  • 157
  • 158
  • 159
  • 160
  • 161
  • 162
  • 163
  • 164
  • 165
  • 166
  • 167
  • 168
  • 169
  • 170
  • 171
  • 172
  • 173
  • 174
  • 175
  • 176
  • 177
  • 178
  • 179
  • 180

Appendix C.
Optimizing Laser Parameters
other frequencies will generally not be absorbed.
The measured transmittance
at the center of the absorption line gives an estimate of the laser multimode
power. Two different methods are used depending on the TGA model
TGA100 and TGA100A
TGA100s and TGA100As have a long sample cell and a short reference cell.
The laser’s multimode power can be estimated by temporarily putting reference
gas in the long sample cell.
This increases the path length by a factor of almost
34.
This test is best performed in conditions in which there is not an excessive
amount of absorption as this can lead to two possible problems.
First, the absorption lines become broader, and absorption in the tails of the
absorption line can reduce the response at the edges of the spectral scan that are
assumed to be 100% transmittance.
This will give an error in the estimate of
multimode power, especially if there is another absorption line nearby.
Second, too much absorption will increase the chances of absorbing the
multimode power in some other absorption lines of the gas.
The correct
absorption can usually be achieved by adjusting the pressure in the sample cell.
To begin, go to the
Settings > Laser > Other
window and set the
Multimode
power %
to zero. Then note the reference transmittance in the
Laser Settings
window.
The
transmittance
measurement
is
affected
by
detector
nonlinearity (see Appendix D.4,
Detector Linearity Coefficients
).
The reference detector is more linear because it has a smaller
signal. Therefore this measurement should be based on the
reference detector, not the sample detector.
To achieve the optimum amount of absorption, first note the reference
transmittance with reference gas in the reference cell only (the normal
configuration). Normally, the reference gas concentration is chosen to give
approximately 50% absorption. For this test, however, it is best to have
between 70% and 80% absorption. Usually this can be accomplished by
reducing the sample flow to reduce the pressure in the analyzer. When the
proper absorption is achieved, record the sample cell pressure.
It is acceptable for the transmittance to be less than 70%, but it
must NOT be greater than 80%.
To configure the instrument for the multimode test, turn off the sample pump,
swap the reference gas and sample connections, and turn the pump back on.
This will put the reference gas in the long sample cell and the air sample in the
short reference cell.
The total flow to the sample pump should be the same as
before, so the pressure should also be the same.
Verify the pressure is within
10% of the previous pressure.
Adjust the laser modulation current as needed – the increased absorption may
make the line wider.
Record the reference transmittance, shown in the
Laser
Settings
window.
This is the estimate of the laser's multimode power. Enter
this value as the
Multimode power (%)
parameter in the
Settings > Laser >
Other
window.
Verify the reference transmittance is near zero.
NOTE
NOTE
C-13