Campbell Scientific TGA200A TGA Series Trace Gas Analzyers - Page 62

Shutdown Procedure, cooled laser, fill the laser dewar

Page 62 highlights

TGA Series Trace Gas Analyzers another absorption line. A persistent trend over time in the DC current may indicate premature aging of the laser. 3. Record the reference transmittance and compare it to previously recorded values. A change in the reference transmittance may indicate a problem with the reference gas supply. It can also be caused by a shift to a different absorption line, or a change in the pressure or temperature in the sample cell. 4. Record the sample pressure and compare it to previously recorded values. If the TGA sample pressure is actively controlled by the datalogger, a change in the pressure indicates a problem with the flows and pressures in the sampling system. If the pressure is not actively controlled it will be determined by the sample flow and the pump capacity. In this case, the pressure will decrease over time as the sample intake filter(s) becomes plugged. 5. If the TGA is equipped with a LN2 cooled laser, record the laser heater voltage and compare it to previously recorded values. The vacuum inside the laser dewar will gradually degrade. This degradation reduces the thermal isolation between the outer wall of the laser dewar and the laser itself. Over time, as more heat is transferred to the laser by the degraded vacuum, less heat is needed to maintain the laser at the set temperature, and the laser heater voltage will gradually decrease. Therefore, monitoring the laser heater voltage may give an indication of when it is time to evacuate the dewar. If the TGA is equipped with a TE-cooled laser, record the laser cooler voltage. 6. Verify that the concentration and concentration noise are as expected. 7. If the TGA is equipped with a LN2-cooled laser, fill the laser dewar with liquid nitrogen as needed. If the TGA is equipped with LN2cooled detectors, fill the detector dewar as needed. 8. Check the reference gas tank and regulator pressure. Check other tanks (air gap purge, calibration, etc.) as needed. Monitor the change in pressure as these gases are consumed, to gauge when to order replacement tanks and to identify possible leaks. 7.1.3 Shutdown Procedure This section describes the routine shutdown procedure for the TGA. 1. If a PC is connected to the TGA, turn off data collection if data is being collected, and quit the TGA program. 2. Shut off the TGA sample pump (and bypass pump, if applicable). 3. Shut off power to the TGA. 4. Shut off reference gas supply. 5. Shut off air gap purge supply, if applicable. 6. Shut off calibration gas supplies, if applicable. 48

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135
  • 136
  • 137
  • 138
  • 139
  • 140
  • 141
  • 142
  • 143
  • 144
  • 145
  • 146
  • 147
  • 148
  • 149
  • 150
  • 151
  • 152
  • 153
  • 154
  • 155
  • 156
  • 157
  • 158
  • 159
  • 160
  • 161
  • 162
  • 163
  • 164
  • 165
  • 166
  • 167
  • 168
  • 169
  • 170
  • 171
  • 172
  • 173
  • 174
  • 175
  • 176
  • 177
  • 178
  • 179
  • 180

TGA Series Trace Gas Analyzers
another absorption line.
A persistent trend over time in the DC
current may indicate premature aging of the laser.
3.
Record the reference transmittance and compare it to previously
recorded values.
A change in the reference transmittance may
indicate a problem with the reference gas supply.
It can also be
caused by a shift to a different absorption line, or a change in the
pressure or temperature in the sample cell.
4.
Record the sample pressure and compare it to previously recorded
values.
If the TGA sample pressure is actively controlled by the
datalogger, a change in the pressure indicates a problem with the
flows and pressures in the sampling system.
If the pressure is not
actively controlled it will be determined by the sample flow and the
pump capacity. In this case, the pressure will decrease over time as
the sample intake filter(s) becomes plugged.
5.
If the TGA is equipped with a LN
2
cooled laser, record the laser
heater voltage and compare it to previously recorded values.
The
vacuum inside the laser dewar will gradually degrade.
This
degradation reduces the thermal isolation between the outer wall of
the laser dewar and the laser itself.
Over time, as more heat is
transferred to the laser by the degraded vacuum, less heat is needed to
maintain the laser at the set temperature, and the laser heater voltage
will gradually decrease. Therefore, monitoring the laser heater
voltage may give an indication of when it is time to evacuate the
dewar. If the TGA is equipped with a TE-cooled laser, record the
laser cooler voltage.
6.
Verify that the concentration and concentration noise are as expected.
7.
If the TGA is equipped with a LN
2
-cooled laser, fill the laser dewar
with liquid nitrogen as needed.
If the TGA is equipped with LN
2
-
cooled detectors, fill the detector dewar as needed.
8.
Check the reference gas tank and regulator pressure.
Check other
tanks (air gap purge, calibration, etc.) as needed.
Monitor the change
in pressure as these gases are consumed, to gauge when to order
replacement tanks and to identify possible leaks.
7.1.3 Shutdown Procedure
This section describes the routine shutdown procedure for the TGA.
1.
If a PC is connected to the TGA, turn off data collection if data is
being collected, and quit the TGA program.
2.
Shut off the TGA sample pump (and bypass pump, if applicable).
3.
Shut off power to the TGA.
4.
Shut off reference gas supply.
5.
Shut off air gap purge supply, if applicable.
6.
Shut off calibration gas supplies, if applicable.
48