Campbell Scientific TGA200A TGA Series Trace Gas Analzyers - Page 60

Run the PC Software, 6.3 Detailed Setup Instructions, 7. Operation, 7.1 Routine Operation

Page 60 highlights

TGA Series Trace Gas Analyzers The identification string is also displayed and edited on the Settings tab of the Device Configuration Utility. The identification string is intended as a device nickname and to be a more user friendly way than the serial number by which to identify which TGA is connected to the PC software. The identification string is displayed at the top of the toolbar when the TGA software is connected to a TGA. The identification string can be any text, up to 19 characters. 6.2.5 Run the PC Software Start the TGA Windows (for LN2-cooled laser) or TGA TEC (for TE-cooled laser) software, which will bring up the toolbar. Descriptions of the toolbar functions are given in greater detail in Section 7.2, Software User Interface. 6.3 Detailed Setup Instructions When the TGA is first installed, or if it is reconfigured (for example, with a new laser) the operational parameters must be set for optimal performance. Appendices A through E give detailed instructions to configure the TGA for a specific gas species, performing the optical alignment, optimizing the laser parameters, optimizing the detector parameters, and calibration. If the TGA has already been configured, see Section 7.1.1, Startup Procedure, for routine startup instructions. 7. Operation 7.1 Routine Operation Once the TGA has been set up, it should be checked periodically to verify proper operation and to perform routine maintenance. Consistently recording operating parameters such as pressure and detector signals is strongly encouraged. Significant changes in these values from one operational period to the next, is often indicative of problems with the system. 7.1.1 Startup Procedure 1. This section describes the routine procedure for starting the TGA. If the TGA is equipped with a LN2 cooled laser, cool the laser dewar. NOTE If the TGA is equipped with a TE-cooled laser, it will be cooled automatically. 2. If the TGA is equipped with LN2-cooled detectors (used for long wavelength operation), cool the detectors with liquid nitrogen. NOTE If the TGA is equipped with the standard TE-cooled detectors, they will be cooled automatically. 3. Start the sample vacuum pump. 4. Turn on the reference gas. A regulator pressure of 0 psig and a flow rate of approximately 10 ml/min are recommended. 46

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135
  • 136
  • 137
  • 138
  • 139
  • 140
  • 141
  • 142
  • 143
  • 144
  • 145
  • 146
  • 147
  • 148
  • 149
  • 150
  • 151
  • 152
  • 153
  • 154
  • 155
  • 156
  • 157
  • 158
  • 159
  • 160
  • 161
  • 162
  • 163
  • 164
  • 165
  • 166
  • 167
  • 168
  • 169
  • 170
  • 171
  • 172
  • 173
  • 174
  • 175
  • 176
  • 177
  • 178
  • 179
  • 180

TGA Series Trace Gas Analyzers
The identification string is also displayed and edited on the
Settings
tab of the
Device Configuration Utility
. The identification string is intended as a device
nickname and to be a more user friendly way than the serial number by which
to identify which TGA is connected to the PC software. The identification
string is displayed at the top of the toolbar when the TGA software is
connected to a TGA. The identification string can be any text, up to 19
characters.
6.2.5 Run the PC Software
Start
the TGA Windows
(for LN
2
-cooled laser) or
TGA TEC
(for TE-cooled
laser) software, which will bring up the toolbar.
Descriptions of the toolbar
functions are given in greater detail in Section 7.2,
Software User Interface
.
6.3
Detailed Setup Instructions
When the TGA is first installed, or if it is reconfigured (for example, with a
new laser) the operational parameters must be set for optimal performance.
Appendices A through E give detailed instructions to configure the TGA for a
specific gas species, performing the optical alignment, optimizing the laser
parameters, optimizing the detector parameters, and calibration. If the TGA
has already been configured, see Section 7.1.1,
Startup Procedure
, for routine
startup instructions.
7.
Operation
7.1
Routine Operation
Once the TGA has been set up, it should be checked periodically to verify
proper operation and to perform routine maintenance. Consistently recording
operating parameters such as pressure and detector signals is strongly
encouraged.
Significant changes in these values from one operational period to
the next, is often indicative of problems with the system.
7.1.1 Startup Procedure
1.
This section describes the routine procedure for starting the TGA.
If
the TGA is equipped with a LN
2
cooled laser, cool the laser dewar.
If the TGA is equipped with a TE-cooled laser, it will be cooled
automatically.
2.
If the TGA is equipped with LN
2
-cooled detectors (used for long
wavelength operation), cool the detectors with liquid nitrogen.
If the TGA is equipped with the standard TE-cooled detectors,
they will be cooled automatically.
3.
Start the sample vacuum pump.
4.
Turn on the reference gas. A regulator pressure of 0 psig and a flow
rate of approximately 10 ml/min are recommended.
NOTE
NOTE
46