Campbell Scientific TGA200A TGA Series Trace Gas Analzyers - Page 57
TGA Series Trace Gas Analyzers, Device Configuration utility Settings Editor tab
View all Campbell Scientific TGA200A manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 57 highlights
TGA Series Trace Gas Analyzers tab to review or edit these settings. The IP address, subnet mask, and gateway should be set as directed by your information technology support group. Set the IP port for the TGA to 3000. FIGURE 6-18. Device Configuration utility Settings Editor tab NOTE The TGA requires a static IP address. Some local area networks are set up to require the network administrator to know the MAC address of devices connected to the network. The Settings Editor tab also displays the TGAs MAC address, but does not allow it to be edited. The MAC address of each TGA is set at the factory and is stored in nonvolatile memory. If the memory storing the MAC address becomes corrupted, it will be displayed as all zeros (0x000...). To reenter the MAC address, execute the following steps: 1. Power down the TGA. 43