114, 115, and 117 Users Manual To open the case for fuse replacement: 1. Remove the test leads from the Meter 2. Remove the Meter from its holster. 3. Remove two screws from the case bottom. 4. Separate the case bottom from the case top. 5. Remove the fuse from its holder and replace it with an 11 A, 1000 V, FAST fuse having a minimum interrupt rating of 17,000 A. Use only Fluke PN 803293. 6. To re-assemble the Meter, first attach the case bottom to the case top, then install the two screws. Finally, insert the Meter into its holster. Cleaning Wipe the case with a damp cloth and mild detergent. Do not use abrasives, isopropyl alcohol, or solvents to clean the case or lens/window. Dirt or moisture in the terminals can affect readings. 18

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26

114, 115, and 117
Users Manual
18
To open the case for fuse replacement:
1.
Remove the test leads from the Meter
2.
Remove the Meter from its holster.
3.
Remove two screws from the case bottom.
4.
Separate the case bottom from the case top.
5.
Remove the fuse from its holder and replace it with an
11 A, 1000 V, FAST fuse having a minimum interrupt
rating of 17,000 A. Use only Fluke PN 803293.
6.
To re-assemble the Meter, first attach the case bottom
to the case top, then install the two screws. Finally,
insert the Meter into its holster.
Cleaning
Wipe the case with a damp cloth and mild detergent.
Do
not use abrasives, isopropyl alcohol, or solvents to clean
the case or lens/window
. Dirt or moisture in the terminals
can affect readings.