Fluke 124B/S Service Manual - Page 36
Input A and B Base Line Jump Test, beeped twice
View all Fluke 124B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 36 highlights
Fluke 123/124/125 Service Manual Figure 3-2. Menu item selection ST7968.WMF If an item is selected, it is marked by . Not selected items are marked by □ If a selected item is highlighted, an then is pressed, the item remains selected. You can also navigate through the menu using . To conform the highlighted item you must press . Before performing the Input A and Input B tests, the test tool must be set in a defined state, by performing a RESET. Proceed as follows to reset the test tool: • Press to turn the test tool off. • Press and hold . • Press and release to turn the test tool on. Wait until the test tool has beeped twice, and then release beeped twice, the RESET was successful. . When the test tool has For most tests, you must turn Input B on. Input A is always on. Proceed as follows to turn Input B on: • Press to open the Meter B menu. • Using select INPUT B: □ ON . • Press to confirm the selection; the □ mark changes to ■ . The active setting from the next item group will be highlighted (for example Vac ), and maintained after leaving the menu. • Press to exit the menu. 3.5.1 Input A and B Base Line Jump Test Proceed as follows to check the Input A and Input B base line jump: 3-6