Fluke 124B/S Service Manual - Page 54
Resistance Measurements Test, Select the following test tool setup
View all Fluke 124B/S manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 54 highlights
Fluke 123/124/125 Service Manual 7. When you are finished, set the 5500A to Standby Sensitivity (Scope) 200V/div 500V/div Table 3-7. V DC and V AC High Voltage Verification Tests Range (Meter) 500V 5500A output Vrms 0V 5500A Frequency DC Main (DC) Reading A-B -000.5 to +000.5 Secondary (AC) Reading A-B 1250V +500V -500V 500V 500V 600V 600V +600V -600V 0V DC DC 60Hz 10 kHz 10 kHz 60Hz DC DC DC +497.0 to +503.0 -497.0 to -503.0 +0.592 to +0.608 -0.592 to -0.608 -0.005 to +0.005 494.0 to 506.0 486.0 to 514.0 0.570 to 0.630 0.584 to 0.616 3.5.16 Resistance Measurements Test Proceed as follows: 1. Connect the test tool to the 5500A as shown in Figure 3-9. 3-24 Figure 3-9. Test Tool Input A to 5500A Normal Output 4-Wire 2. Select the following test tool setup: • Press to select auto ranging (AUTO in top of display). ST8003.WMF