Resistance Measurements Test, Select the following test tool setup

Page 54 highlights

Fluke 123/124/125 Service Manual 7. When you are finished, set the 5500A to Standby Sensitivity (Scope) 200V/div 500V/div Table 3-7. V DC and V AC High Voltage Verification Tests Range (Meter) 500V 5500A output Vrms 0V 5500A Frequency DC Main (DC) Reading A-B -000.5 to +000.5 Secondary (AC) Reading A-B 1250V +500V -500V 500V 500V 600V 600V +600V -600V 0V DC DC 60Hz 10 kHz 10 kHz 60Hz DC DC DC +497.0 to +503.0 -497.0 to -503.0 +0.592 to +0.608 -0.592 to -0.608 -0.005 to +0.005 494.0 to 506.0 486.0 to 514.0 0.570 to 0.630 0.584 to 0.616 3.5.16 Resistance Measurements Test Proceed as follows: 1. Connect the test tool to the 5500A as shown in Figure 3-9. 3-24 Figure 3-9. Test Tool Input A to 5500A Normal Output 4-Wire 2. Select the following test tool setup: • Press to select auto ranging (AUTO in top of display). ST8003.WMF

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96

Fluke 123/124/125
Service Manual
3-24
7.
When you are finished, set the 5500A
to Standby
Table 3-7. V DC and V AC High Voltage Verification Tests
Sensitivity
(Scope)
Range
(Meter)
5500A
output Vrms
5500A
Frequency
Main (DC)
Reading A-B
Secondary (AC)
Reading A-B
200V/div
500V
0V
DC
-000.5 to
+000.5
+500V
DC
+497.0 to
+503.0
-500V
DC
-497.0 to -503.0
500V
60Hz
494.0 to 506.0
500V
10 kHz
486.0 to 514.0
500V/div
1250V
600V
10 kHz
0.570 to 0.630
600V
60Hz
0.584 to 0.616
+600V
DC
+0.592 to
+0.608
-600V
DC
-0.592 to -0.608
0V
DC
-0.005 to
+0.005
3.5.16 Resistance Measurements Test
Proceed as follows:
1.
Connect the test tool to the 5500A as shown in Figure 3-9.
ST8003.WMF
Figure 3-9. Test Tool Input A to 5500A Normal Output 4-Wire
2.
Select the following test tool setup:
Press
to select auto ranging (
AUTO
in top of display).