Fluke 124B/S Service Manual - Page 38

Input A Trigger Sensitivity Test, output, MODE levsine. Set the 5500A to Operate OPR.

Page 38 highlights

Fluke 123/124/125 Service Manual Observe the Input B trace, and check to see if it is set to the same position after changing the sensitivity. The allowed difference is ±0.04 division (= 1 pixel). 8. When you are finished, remove the Input A and Input B short. 3.5.2 Input A Trigger Sensitivity Test Proceed as follows to test the Input A trigger sensitivity: 1. Connect the test tool to the 5500A as shown in Figure 3-3. Figure 3-3. Test Tool Input A to 5500A Scope Output 50Ω ST8004.WMF 2. Select the following test tool setup: • Press to select auto ranging (AUTO in top of display). Do not press anymore! • Using change the sensitivity to select manual sensitivity ranging, and lock the Input A sensitivity on 200 mV/div. 3. Set the 5500A to source a 5 MHz leveled sine wave of 100 mV peak-to-peak (SCOPE output, MODE levsine). Set the 5500A to Operate (OPR). 4. Adjust the amplitude of the sine wave to 0.5 division on the display. 5. Verify that the signal is well triggered. If it is not, press to enable the up/down arrow keys for Trigger Level adjustment; adjust the trigger level using and verify that the signal will be triggered now. The trigger level is indicated by the trigger icon ( ). 6. Set the 5500A to source a 25 MHz (Fluke 123) or 40 MHz (Fluke 124/125) leveled sine wave of 400 mV peak-to-peak. 7. Adjust the amplitude of the sine wave to 1.5 divisions on the test tool display. 8. Verify that the signal is well triggered. If it is not, press to enable the up/down arrow keys for Trigger Level adjustment; adjust the trigger level and verify that the signal will be triggered now. 3-8

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96

Fluke 123/124/125
Service Manual
3-8
Observe the Input B trace, and check to see if it is set to the same position after changing
the sensitivity.
The allowed difference is
±
0.04 division (= 1 pixel).
8.
When you are finished, remove the Input A and Input B short.
3.5.2 Input A Trigger Sensitivity Test
Proceed as follows to test the Input A trigger sensitivity:
1.
Connect the test tool to the 5500A as shown in Figure 3-3.
ST8004.WMF
Figure 3-3. Test Tool Input A to 5500A Scope Output 50
Ω
2.
Select the following test tool setup:
Press
to select auto ranging (
AUTO
in top of display).
Do not press
anymore!
Using
change the sensitivity to select manual sensitivity ranging, and lock
the Input A sensitivity on 200 mV/div.
3.
Set the 5500A to source a 5 MHz leveled sine wave of 100 mV peak-to-peak (SCOPE
output, MODE levsine). Set the 5500A to Operate (OPR).
4.
Adjust the amplitude of the sine wave to 0.5 division on the display.
5.
Verify that the signal is well triggered.
If it is not, press
to enable the up/down arrow keys for Trigger Level adjustment;
adjust the trigger level using
and verify that the signal will be triggered now.
The trigger level is indicated by the trigger icon (
).
6.
Set the 5500A to source a 25 MHz (Fluke 123) or 40 MHz (Fluke 124/125) leveled sine
wave of 400 mV peak-to-peak.
7.
Adjust the amplitude of the sine wave to 1.5 divisions on the test tool display.
8.
Verify that the signal is well triggered.
If it is not, press
to enable the up/down arrow keys for Trigger Level adjustment;
adjust the trigger level and verify that the signal will be triggered now.