Fluke 1736/B Datasheet - Page 5

Power supply, Data acquisition, Basic interval, Demand Interval Energy Meter Mode, Power quality

Page 5 highlights

Electrical specifications Power supply Voltage range 100 V to 500 V using safety plug input when powering from the measurement circuit 100 V to 240 V using standard power cord (IEC 60320 C7) Power consumption Maximum 50 VA (max. 15 VA when powered using IEC 60320 input) Efficiency ≥ 68.2 % (in accordance with energy efficiency regulations) Maximum no-load consumption < 0.3 W only when powered using IEC 60320 input Mains power frequency 50/60 Hz ± 15 % Battery Li-ion 3.7 V, 9.25 Wh, customer-replaceable On-battery runtime Four hours in standard operating mode, up to 5.5 hours in power saving mode Charging time < 6 hours Data acquisition Resolution 16-bit synchronous sampling Sampling frequency 10.24 kHz at 50/60 Hz, synchronized to mains frequency Input signal frequency 50/60 Hz (42.5 to 69 Hz) Circuit types 1-ϕ, 1-ϕ IT, split phase, 3-ϕ delta, 3-ϕ wye, 3-ϕ wye IT, 3-ϕ wye balanced, 3-ϕ Aron/Blondel (2-element delta), 3-ϕ delta open leg, currents only (load studies) Data storage Memory size Internal flash memory (not user replaceable) Typical 10 logging sessions of 8 weeks with 1-minute intervals and 500 events1 Basic interval Measured parameters Voltage, current, aux, frequency, THD V, THD A, power, power factor, fundamental power, DPF, energy Averaging interval User selectable: 1 sec, 5 sec, 10 sec, 30 sec, 1 min, 5 min, 10 min, 15 min, 30 min Voltage, Current: Full cycle RMS updated every half cycle (URMS1/2 according to Averaging time min/max values IEC61000-4-30 Aux, Power: 200ms Demand Interval (Energy Meter Mode) Measured parameters Energy (Wh, varh, VAh), PF, maximum demand, cost of energy Interval User selectable: 5 min, 10 min, 15 min, 20 min, 30 min, off Power quality measurements Measured parameter Voltage, frequency, unbalance, voltage harmonics, THD V, current, harmonics, THD A, TDD Averaging interval 10 min Individual harmonics 2nd ...50th harmonic Total harmonic distortion Calculated on 50 harmonics Events Voltage: dips, swells, interruptions, current: inrush current Triggered recordings Full cycle RMS updated every half cycle of voltage and current (Urms1/2 according to IEC61000-4-30) Waveform of voltage and current 1The number of possible logging sessions and logging period depends on user requirements. MyFlukeStore www. 5SFhluokpe CfoorproFraltuiokne pFlruokde 1u7c3t6sanodn1l7i3n8eTharte:e-Phase Power Loggers .com 1.877.766.5412

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10

Electrical specifications
Power supply
Voltage range
100 V to 500 V using safety plug input when powering from the
measurement circuit
100 V to 240 V using standard power cord (IEC 60320 C7)
Power consumption
Maximum 50 VA (max. 15 VA when powered using IEC 60320 input)
Efficiency
≥ 68.2 % (in accordance with energy efficiency regulations)
Maximum no-load consumption
< 0.3 W only when powered using IEC 60320 input
Mains power frequency
50/60 Hz ± 15 %
Battery
Li-ion 3.7 V, 9.25 Wh, customer-replaceable
On-battery runtime
Four hours in standard operating mode, up to 5.5 hours in power saving mode
Charging time
< 6 hours
Data acquisition
Resolution
16-bit synchronous sampling
Sampling frequency
10.24 kHz at 50/60 Hz, synchronized to mains frequency
Input signal frequency
50/60 Hz (42.5 to 69 Hz)
Circuit types
1-
ϕ
, 1-
ϕ
IT, split phase, 3-
ϕ
delta, 3-
ϕ
wye, 3-
ϕ
wye IT, 3-
ϕ
wye balanced, 3-
ϕ
Aron/Blondel (2-element delta), 3-
ϕ
delta open leg, currents only (load studies)
Data storage
Internal flash memory (not user replaceable)
Memory size
Typical 10 logging sessions of 8 weeks with 1-minute intervals and 500 events
1
Basic interval
Measured parameters
Voltage, current, aux, frequency, THD V, THD A, power, power factor, fundamental
power, DPF, energy
Averaging interval
User selectable: 1 sec, 5 sec, 10 sec, 30 sec, 1 min, 5 min, 10 min, 15 min, 30 min
Averaging time min/max values
Voltage, Current: Full cycle RMS updated every half cycle (URMS1/2 according to
IEC61000-4-30
Aux, Power: 200ms
Demand Interval (Energy Meter Mode)
Measured parameters
Energy (Wh, varh, VAh), PF, maximum demand, cost of energy
Interval
User selectable: 5 min, 10 min, 15 min, 20 min, 30 min, off
Power quality measurements
Measured parameter
Voltage, frequency, unbalance, voltage harmonics, THD V, current, harmonics, THD
A, TDD
Averaging interval
10 min
Individual harmonics
2nd ...50th harmonic
Total harmonic distortion
Calculated on 50 harmonics
Events
Voltage: dips, swells, interruptions, current: inrush current
Triggered recordings
Full cycle RMS updated every half cycle of voltage and current (Urms1/2 according
to IEC61000-4-30)
Waveform of voltage and current
1
The number of possible logging sessions and logging period depends on user requirements.