Fluke 1736/B Datasheet - Page 7

m up to 4000 m derate to 1000 V CAT II/600 V CAT III/300 V CAT IV

Page 7 highlights

Environmental specifications Operating temperature Storage temperature Operating humidity Operating altitude Storage altitude Enclosure Vibration Safety Electromagnetic compatibility (EMC) Temperature coefficient General specifications Color LCD display Warranty Dimensions Weight Tamper protection -10 °C to +50 °C (14 °F to 122 °F) -20 °C to +60 °C (−4 °F to 140 °F), with battery: -20 °C to +50 °C (−4 °F to 122 °F) 10 °C to 30 °C (50 °F to 86 °F) max. 95 % RH 30 °C to 40 °C (86 °F to 104 °F) max. 75 % RH 40 °C to 50 °C (104 °F to 122 °F) max. 45 % RH 2000 m (up to 4000 m derate to 1000 V CAT II/600 V CAT III/300 V CAT IV) 12,000 m IP50 in accordance with EN60529 MIL-T-28800E, Type 3, Class III, Style B IEC 61010-1 IEC Mains Input: Overvoltage Category II, Pollution Degree 2 Voltage Terminals: Overvoltage Category IV, Pollution Degree 2 IEC 61010-2-031: CAT IV 600 V / CAT III 1000 V EN 61326-1: Industrial CISPR 11: Group 1, Class A Korea (KCC): Class A Equipment (industrial broadcasting and communication equipment) USA (FCC): 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103 0.1 x accuracy specification/°C 4.3-inch active matrix TFT, 480 pixels x 272 pixels, resistive touch panel Instrument and power supply: Two-years (battery not included) Accessories: one-year Calibration cycle: two-years Instrument: 19.8 cm x 16.7 cm x 5.5 cm (7.8 in x 6.6 in x 2.2 in) Power supply: 13.0 cm x 13.0 cm x 4.5 cm (5.1 in x 5.1 in x 1.8 in) Instrument with power supply attached: 19.8 cm x 16.7 cm x 9 cm (7.8 in x 6.6 in x 3.5 in) Instrument: 1.1 kg (2.5 lb) Power supply: 400 g (0.9 lb) Kensington lock slot MyFlukeStore www. 7SFhluokpe CfoorproFraltuiokne pFlruokde 1u7c3t6sanodn1l7i3n8eTharte:e-Phase Power Loggers .com 1.877.766.5412

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10

Environmental specifications
Operating temperature
-10 °C to +50 °C (14 °F to 122 °F)
Storage temperature
-20 °C to +60 °C (−4 °F to 140 °F), with battery: -20 °C to +50 °C (−4 °F to 122 °F)
Operating humidity
10 °C to 30 °C (50 °F to 86 °F) max. 95 % RH
30 °C to 40 °C (86 °F to 104 °F) max. 75 % RH
40 °C to 50 °C (104 °F to 122 °F) max. 45 % RH
Operating altitude
2000 m (up to 4000 m derate to 1000 V CAT II/600 V CAT III/300 V CAT IV)
Storage altitude
12,000 m
Enclosure
IP50 in accordance with EN60529
Vibration
MIL-T-28800E, Type 3, Class III, Style B
Safety
IEC 61010-1
IEC Mains Input: Overvoltage Category II, Pollution Degree 2
Voltage Terminals: Overvoltage Category IV, Pollution Degree 2
IEC 61010-2-031: CAT IV 600 V / CAT III 1000 V
Electromagnetic compatibility
(EMC)
EN 61326-1: Industrial CISPR 11: Group 1, Class A
Korea (KCC): Class A Equipment (industrial broadcasting and
communication equipment)
USA (FCC): 47 CFR 15 subpart B. This product is considered an exempt device per
clause 15.103
Temperature coefficient
0.1 x accuracy specification/°C
General specifications
Color LCD display
4.3-inch active matrix TFT, 480 pixels x 272 pixels, resistive touch panel
Warranty
Instrument and power supply: Two-years (battery not included)
Accessories: one-year
Calibration cycle: two-years
Dimensions
Instrument: 19.8 cm x 16.7 cm x 5.5 cm (7.8 in x 6.6 in x 2.2 in)
Power supply: 13.0 cm x 13.0 cm x 4.5 cm (5.1 in x 5.1 in x 1.8 in)
Instrument with power supply attached: 19.8 cm x 16.7 cm x 9 cm
(7.8 in x 6.6 in x 3.5 in)
Weight
Instrument: 1.1 kg (2.5 lb)
Power supply: 400 g (0.9 lb)
Tamper protection
Kensington lock slot
Shop for Fluke products online at:
1.877.766.5412
www.
MyFlukeStore