The cables listed in Table 3 are necessary for verification and calibration adjustment.
View all Fluke 1750 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 15 highlights
Power Recorder Required Equipment The cables listed in Table 3 are necessary for verification and calibration adjustment. To assemble the cables, see cable assembly figures in the Test Cables section. Cable Table 3. Required Cables Required Characteristics Notes Ethernet cable CAT 5 with RJ-45 connectors Supplied with the Recorder Voltage calibration cable with BNC-banana adapter Double banana to safety banana and spade lug See Figure 5 Four jumper cables, safetybanana type Current calibration cable Rogowski resistor measurement cable Shielded cable with double banana to CT pins in Redel connector with 100 A CT ID resistor Safety banana to Redel See Figure 6 See Figure 4 Rogowski calibration cable Shielded cable with double banana to Rogowski pins in Redel connector with 100 A Rogowski ID resistor See Figure 3 Ground calibration cable Lug to banana See Figure 2 Fluke recommends the use of Pomona brand cables and banana plugs when building these cables. 7