Fluke 1750 Fluke 1750 Calibration Manual - Page 22

Calibration, Required Equipment, Calibration Adjustment - case

Page 22 highlights

1750 Calibration Manual Applie d Voltage [Vrms] 800 850 900 950 1000 Expected Reading [Vpk] 1131 1202 1273 1344 1414 Table 8. Impulse Verification Inputs Channel Measured Magnitude |Vpk| AN (A) BN (B) CN (C) NG (N) NG (G) Limits [Vpk] ± 77 ± 80 ± 84 ± 87 ± 91 Calibration Required Equipment The required equipment and cables for calibrating the Recorder are listed in Tables 2 and 3. XWWarning To avoid electrical shock, personal injury, or fire: • Do not perform the calibration procedures or calibration verification tests described in this manual unless you are qualified to do so. • Repairs or servicing should be performed only by qualified personnel. Calibration Adjustment The Recorder features closed-case calibration adjustment using known reference sources. During calibration, the Recorder measures the applied reference source, calculates correction factors, and stores the correction factors in nonvolatile memory. Perform calibration adjustment if the Recorder fails any performance test in the verification procedure. Note Calibration should be performed under ambient temperature conditions between 20 °C and 30 °C. Calibration consists of four parts: I-Wave, I-Rogowski, V-Wave, and V-Impulse. Each part calibrates one measurement system in the Recorder. For each part, the calibration values are checked to be sure the Recorder is working correctly and calibration results are valid. All of the calibration steps for each of the four measurement systems must be completed before the calibration values can be written to the Recorder's nonvolatile memory. I-Wave, I-Rogowski, and V-Wave channels are calibrated using 55 Hz sine wave signals to minimize 50 Hz or 60 Hz power line interference with the calibration measurements. V-Impulse channels are calibrated using dc voltages, averaging calibration measurements to suppress power line interference. 14

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28

1750
Calibration Manual
14
Table 8. Impulse Verification Inputs
Applie
d
Voltage
[Vrms]
Expected
Reading
[Vpk]
Channel
Measured Magnitude
|Vpk|
Limits
[Vpk]
800
1131
AN (A)
± 77
850
1202
BN (B)
± 80
900
1273
CN (C)
± 84
950
1344
NG (N)
± 87
1000
1414
NG (G)
± 91
Calibration
Required Equipment
The required equipment and cables for calibrating the Recorder are listed in Tables 2 and
3.
XW
Warning
To avoid electrical shock, personal injury, or fire:
Do not perform the calibration procedures or calibration
verification tests described in this manual unless you are
qualified to do so.
Repairs or servicing should be performed only by qualified
personnel.
Calibration Adjustment
The Recorder features closed-case calibration adjustment using known reference sources.
During calibration, the Recorder measures the applied reference source, calculates
correction factors, and stores the correction factors in nonvolatile memory.
Perform calibration adjustment if the Recorder fails any performance test in the
verification procedure.
Note
Calibration should be performed under ambient temperature conditions
between 20
°
C and 30
°
C.
Calibration consists of four parts: I-Wave, I-Rogowski, V-Wave, and V-Impulse. Each
part calibrates one measurement system in the Recorder. For each part, the calibration
values are checked to be sure the Recorder is working correctly and calibration results are
valid. All of the calibration steps for each of the four measurement systems must be
completed before the calibration values can be written to the Recorder’s nonvolatile
memory.
I-Wave, I-Rogowski, and V-Wave channels are calibrated using 55 Hz sine wave signals
to minimize 50 Hz or 60 Hz power line interference with the calibration measurements.
V-Impulse channels are calibrated using dc voltages, averaging calibration measurements
to suppress power line interference.