Fluke 1750 Fluke 1750 Calibration Manual - Page 17
Power View Software Live View, Table 4. AC Voltage Accuracy Verification, Applied - three phase power recorder
View all Fluke 1750 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 17 highlights
Power Recorder Verification Tests Figure 1. Power View Software Live View power analyze live view screen.bmp 7. Connect the calibrator NORMAL HI to inputs A, B, C, and J of the Recorder. 8. Connect the NORMAL LO to input N and the protective conductor terminal . of the Recorder. 9. Apply the voltages from Table 4 and list the Recorder readings for channels AN, BN, CN, and NG in the appropriate areas. Sine waves will be displayed for AN, BN and CN, all at the same amplitude and phase shift. NG will be inverted. Note If the voltage goes over scale, right click on the voltage scale on the left margin of the graphical display, then click on the Zoom to Fit box to resize. Applied 60 Vrms @ Local Hz 240 Vrms @ Local Hz 600 Vrms @ Local Hz Table 4. AC Voltage Accuracy Verification AN Reading BN Reading CN Reading NG Reading Limits ± 1.12 V ± 1.48 V ± 2.20 V 10. Verify that A, B, C, and J LEDs are on. N should be off. 11. Set the calibrator to STBY. 12. Verify that A, B, C, and J LEDs are off. 9