Fluke 1750 Fluke 1750 Calibration Manual - Page 17

Power View Software Live View, Table 4. AC Voltage Accuracy Verification, Applied - three phase power recorder

Page 17 highlights

Power Recorder Verification Tests Figure 1. Power View Software Live View power analyze live view screen.bmp 7. Connect the calibrator NORMAL HI to inputs A, B, C, and J of the Recorder. 8. Connect the NORMAL LO to input N and the protective conductor terminal . of the Recorder. 9. Apply the voltages from Table 4 and list the Recorder readings for channels AN, BN, CN, and NG in the appropriate areas. Sine waves will be displayed for AN, BN and CN, all at the same amplitude and phase shift. NG will be inverted. Note If the voltage goes over scale, right click on the voltage scale on the left margin of the graphical display, then click on the Zoom to Fit box to resize. Applied 60 Vrms @ Local Hz 240 Vrms @ Local Hz 600 Vrms @ Local Hz Table 4. AC Voltage Accuracy Verification AN Reading BN Reading CN Reading NG Reading Limits ± 1.12 V ± 1.48 V ± 2.20 V 10. Verify that A, B, C, and J LEDs are on. N should be off. 11. Set the calibrator to STBY. 12. Verify that A, B, C, and J LEDs are off. 9

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28

Power Recorder
Verification Tests
9
power analyze live view screen.bmp
Figure 1. Power View Software Live View
7.
Connect the calibrator
NORMAL HI
to inputs
A
,
B
,
C,
and
J
of the Recorder.
8.
Connect the
NORMAL LO
to input
N
and the protective conductor terminal
.
of
the Recorder.
9.
Apply the voltages from Table 4 and list the Recorder readings for channels
AN
,
BN,
CN,
and
NG
in the appropriate areas. Sine waves will be displayed for
AN
,
BN
and
CN
, all at the same amplitude and phase shift.
NG
will be inverted.
Note
If the voltage goes over scale, right click on the voltage scale on the left
margin of the graphical display, then click on the
Zoom to Fit
box to
resize.
Table 4. AC Voltage Accuracy Verification
Applied
AN Reading
BN Reading
CN Reading
NG Reading
Limits
60 Vrms
@
Local Hz
± 1.12 V
240 Vrms
@
Local Hz
± 1.48 V
600 Vrms
@
Local Hz
± 2.20 V
10.
Verify that
A
,
B
,
C,
and
J
LEDs are on.
N
should be off.
11.
Set the calibrator to
STBY.
12.
Verify that
A
,
B
,
C,
and
J
LEDs are off.