Fluke 192B - 196B/C - 199B/C Users Manual Amperes (AMP) With Optional Current Probe or Current Shunt Ranges same as VDC, VAC, VAC+DC Probe Sensitivity 100 µV/A, 1 mV/A, 10 mV/A, 100 mV/A, 1 V/A, 10 V/A, and 100 V/A Accuracy same as VDC, VAC, VAC+DC (add current probe or current shunt accuracy) Peak Modes Max peak, Min peak, or pk-to-pk Maximum Voltage with 10:1 probe 1000 V direct (1:1 300 V Maximum Resolution with 10:1 probe 10 mV direct (1:1 1 mV Full Scale Reading 800 counts Accuracy Max peak or Min peak 0.2 division Peak-to-peak 0.4 division Frequency (Hz) Range 1.000 Hz to full bandwidth Full Scale Reading 9 999 counts with at least 10 waveform periods on screen. Accuracy 1 Hz to full bandwidth 0.5 % +2 counts) Duty Cycle (DUTY) Range 4.0 % to 98.0 % Pulse Width (PULSE) Resolution (with GLITCH off 1/100 division Full Scale Reading 999 counts Accuracy 1 Hz to full bandwidth 0.5 % +2 counts) Vpwm (C versions only) Purpose to measure on pulse width modulated signals, like motor drive inverter outputs Principle ......... readings show the effective voltage based on the average value of samples over a whole number of periods of the fundamental frequency Accuracy as Vrms for sinewave signals 100