Fluke 196C FE 192,196,199 C Users Manual - Page 110
Meter, DMM Measurements on Meter Inputs, Meter Input, Meter Functions, General, Continuity CONT
View all Fluke 196C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 110 highlights
Fluke 192B - 196B/C - 199B/C Users Manual Meter Meter Input Input Coupling DC Frequency Response DC to 10 kHz (-3 dB) Input Impedance 1 MΩ (±1 %)//10 pF (±1.5 pF) Max. Input Voltage 1000 V CAT II 600 V CAT III (For detailed specifications, see "Safety") Meter Functions Ranging Auto, Manual Modes Normal, Relative 102 DMM Measurements on Meter Inputs The accuracy of all measurements is within ± (% of reading + number of counts) from 18 °C to 28 °C. Add 0.1x (specific accuracy) for each °C below 18 °C or above 28 °C. General DC Common Mode Rejection (CMRR 100 dB AC Common Mode Rejection at 50, 60, or 400 Hz ......>60 dB Ohms (Ω) Ranges 500.0 Ω, 5.000 kΩ, 50.00 kΩ, 500.0 kΩ, 5.000 MΩ, 30.00 MΩ Full Scale Reading 500 Ω to 5 M 5000 counts 30 M 3000 counts Accuracy 0.6 % +5 counts) Measurement Current 0.5 mA to 50 nA, ±20 % decreases with increasing ranges Open Circuit Voltage 4 V Continuity (CONT) Beep 50 Ω (±30 Ω) Measurement Current 0.5 mA, ±20 % Detection of shorts of 1 ms