Fluke 196C FE 192,196,199 C Users Manual - Page 112
Recorder, Amperes AMP, TrendPlot Meter or Scope, Scope Record
View all Fluke 196C manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 112 highlights
Fluke 192B - 196B/C - 199B/C Users Manual Amperes (AMP) With Optional Current Probe or Current Shunt Ranges same as VDC, VAC, VAC+DC Probe Sensitivity 100 µV/A, 1 mV/A, 10 mV/A, 100 mV/A, 1 V/A, 10 V/A, and 100 V/A Accuracy same as VDC, VAC, VAC+DC (add current probe or current shunt accuracy) Recorder TrendPlot (Meter or Scope) Chart recorder that plots a graph of min and max values of Meter or Scope measurements over time. Measurement Speed 5 measurements/s max Time/Div 5 s/div to 30 min/div Record Size 18000 points Recorded Time Span... 60 min to 22 days(single reading) 30 min to 11 days (dual reading) Time Reference time from start, time of day Scope Record Records scope waveforms in deep memory while displaying the waveform in Roll mode. Source Input A, Input B Max. Sample Speed (5 ms/div to 1 min/div) .........20 MS/s Glitch capture (5 ms/div to 1 min/div 50 ns Time/Div in normal mode 5 ms/div to 2 min/div Record Size 27000 points per input Recorded Time Span 6 s to 48 hours Acquisition Modes Single Sweep Continuous Roll External Triggering Time Reference time from start, time of day 104