Fluke 718-300G Fluke 718 Users Manual - Page 14
Getting Acquainted with the Calibrator, Power Saver - calibration manual
View all Fluke 718-300G manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 14 highlights
718 Series Users Manual Getting Acquainted with the Calibrator Press O to turn the Calibrator on and off. The Calibrator displays pressure and current measurements simultaneously. See Figure 2. The upper part of the display shows the applied pressure or vacuum. (Vacuum is shown as a negative value.) Press U to select a different unit. When you cycle the power off and on, the Calibrator retains the unit you last used. The lower part of the display shows the current (up to 24 mA) applied to the current (mA) inputs. To source loop voltage, press U while pressing O on. Pushbutton operation is described in Table 4. Pump features are shown in Figure 3 and described in Table 5. Power Saver The Calibrator automatically turns off after 30 minutes of inactivity. To reduce this time or disable this feature: 1. With the Calibrator OFF, press O. 2. P.S. xx is displayed, where xx is the turn-off time in minutes. OFF means the power saver is disabled. 3. Press h (W) to decrease or a(X) to increase the turn-off time. 4. To disable, press h until the display shows OFF. The Calibrator resumes normal operation after 2 seconds. 6 Pressure measurement 718 PRCEASLISBURRAETOR HOLD Current mA measurement On/Off Button UNITS LPOOOWPER ZERO DAMP ENTER MMINAX BAmMROAOMDEETHROICLADDJ. SWTITECSHT 0-3R0pAsNi G0E-206kPa 3M0VAX Pressure module input mA COM RANGMEAX-1-75OPOS1I7PkS3PI4.5 P Pressure sensor input (Install filter here) Current input Figure 2. Front Panel Features wh005f.eps