Fluke 718-300G Fluke 718 Users Manual - Page 17
Zeroing with Absolute Pressure Modules, Pump Features
View all Fluke 718-300G manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 17 highlights
Zeroing with Absolute Pressure Modules For zeroing, adjust the Calibrator to read a known pressure. This can be barometric pressure, if it is accurately known, for all but the 700PA3 module. An accurate pressure standard can also apply a pressure within range for any Absolute Pressure Module. Adjust the Calibrator reading as follows: 1. Press and hold Z . 2. Press a(Z) to increase or h (Y) to decrease the Calibrator reading to equal the applied pressure. 3. Release Z to exit the zeroing procedure. Press the U button to convert to any convenient measurement display unit. Pressure Calibrator Zeroing with Absolute Pressure Modules ZERO MMINAX BAmMRAOODMEETHROICLADDJ. SWTITECSHT 0-3R0pAsNi G0E-206kPa mA 3M0VAX COM RANGMEAX-75OPOS1I7PkSPI P Pressure/ Vacuum Switch Pressure/Vacuum Release Control Fine Adjustment Knob Internal Pump Figure 3. Pump Features wh009f.eps 9