Fluke 718-300G Fluke 718 Users Manual - Page 24

Using an External Pump, release of pressure, do not connect the internal - pressure calibrator manual

Page 24 highlights

718 Series Users Manual Using an External Pump XW Warning To avoid damage to the Calibrator and possible release of pressure, do not connect the internal sensor to an external pressure source that exceeds the maximum rated pressure. To develop higher pressure or vacuum, use an external pump (such as the Fluke Model 700PTP). Use a Fluke Pressure Module connected to the pressure module input on the Calibrator. Pressure modules are listed in Table . Make overall connections as shown in Figure 6. Refer to setup and operating instructions included with the pressure module and pump. 16